Literature DB >> 15834914

An MRI-compatible semiautomated vacuum-assisted breast biopsy system: initial feasibility study.

Bruce L Daniel1, Lynetta J Freeman, Jessica M Pyzoha, Terry D McCoy, Robyn L Birdwell, Donna M Bouley, Brennan Movius, John A Hibner.   

Abstract

PURPOSE: To develop an MR-compatible vacuum-assisted core needle breast biopsy system.
MATERIALS AND METHODS: A vacuum-assisted breast biopsy system (Mammotome Hand Held; Ethicon Endo-Surgery, USA) was modified for freehand MRI-guided biopsy in an open, interventional 0.5-T scanner (Signa SP; GE, USA). Probes (11 gauge [G]) were fabricated without significant susceptibility artifact. These mate with an electromechanical hand piece and control system that were modified for use within the MRI scanner. A total of 62 breast lesions were simulated in the mammary tissues of six recently lactating sows by injecting between 0.1 and 1.0 mL of an aqueous gel containing dilute gadopentetate dimeglumine (Gd-DTPA) that formed a bright mass on T1-weighted imaging.
RESULTS: Mechanical performance was satisfactory. Magnetic susceptibility and radiofrequency (RF) artifacts from the 11-G probe were negligible. T1-weighted fast spin echo (FSE) was used to guide biopsy. Up to eight samples were removed per lesion. Overall, 461 samples were obtained in 493 attempts (94%). Sample weights averaged 54 mg (N = 493) compared to 4.6 mg (N = 24) from 14-G titanium MRI-compatible needles. On average, 59% of the attempted samples yielded target lesion material.
CONCLUSION: Preliminary results demonstrate the feasibility of a vacuum-assisted breast biopsy system in the MRI environment. Small 0.1-mL lesions can be biopsied without needle artifacts obscuring the target.

Entities:  

Mesh:

Year:  2005        PMID: 15834914     DOI: 10.1002/jmri.20302

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  6 in total

1.  Modeling of active shimming of metallic needles for interventional MRI.

Authors:  Saikat Sengupta
Journal:  Magn Reson Med       Date:  2020-06-29       Impact factor: 4.668

2.  Study of the performance of a novel 1 mm resolution dual-panel PET camera design dedicated to breast cancer imaging using Monte Carlo simulation.

Authors:  Jin Zhang; Peter D Olcott; Garry Chinn; Angela M K Foudray; Craig S Levine
Journal:  Med Phys       Date:  2007-02       Impact factor: 4.071

3.  Minimal artifact actively shimmed metallic needles in MRI.

Authors:  Saikat Sengupta; Xinqiang Yan; Tamarya L Hoyt; Gary Drake; Anthony Gunderman; Yue Chen
Journal:  Magn Reson Med       Date:  2021-08-19       Impact factor: 4.668

Review 4.  [Interventional MRI of the breast. Indications, technique, results and perspectives].

Authors:  K Hellerhoff; T Schlossbauer; P Herzog; M Reiser
Journal:  Radiologe       Date:  2008-04       Impact factor: 0.635

Review 5.  Status quo and development trend of breast biopsy technology.

Authors:  Yan-Jun Zhang; Lichun Wei; Jie Li; Yi-Qiong Zheng; Xi-Ru Li
Journal:  Gland Surg       Date:  2013-02

Review 6.  [Magnetic resonance imaging in preoperative staging for breast cancer: pros and contras].

Authors:  C K Kuhl; M Braun
Journal:  Radiologe       Date:  2008-04       Impact factor: 0.635

  6 in total