A comparative study of the action of dexamethasone sodium phosphate and dexamethasone acetate in steroid-eluting pacemaker leads.

Suresh Singarayar1, Peter M Kistler, Chantal De Winter, Harry Mond.   

Abstract

BACKGROUND: The aim of this study was to characterize acute and medium-term pacemaker lead performance with the two most commonly used glucocorticosteroids: dexamethasone sodium phosphate and dexamethasone acetate.
METHODS: Forty sets of atrial and ventricular passive-fixation leads containing either dexamethasone sodium phosphate or dexamethasone acetate were implanted as dual chamber pacemakers. Randomization was equally distributed to both arms of the study. Stimulation thresholds, lead impedance, and sensing were measured on the day of implant, day 1, 1 month, 3 months, and 6 months following the implant.
RESULTS: For the dexamethasone sodium phosphate arm, the atrial stimulation thresholds were 0.9 +/- 0.1 V at implant and 0.8 +/- 0.1 V at 6 months, and in the ventricle 0.5 +/- 0.1 V at implant and 0.6 +/- 0.1 V at 6 months. In the dexamethasone acetate arm, the atrial stimulation thresholds were 0.7 +/- 0.1 V at implant and at 6 months, and in the ventricle 0.5 +/- 0.1 V at implant and at 6 months. There were no significant differences between dexamethasone sodium phosphate or dexamethasone acetate leads for stimulation thresholds at any of the intervals of follow-up. P- and R-wave sensing were similarly maintained over the duration of follow-up with no significant differences between groups at any of the intervals of follow-up. Pacing lead impedance showed a trend towards lower values in the dexamethasone acetate arm, which only reached statistical significance at 3 months and beyond for ventricular leads.
CONCLUSIONS: Leads containing dexamethasone sodium phosphate and dexamethasone acetate demonstrate equivalent and excellent acute and medium-term pacemaker lead performance characteristics.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15826265     DOI: 10.1111/j.1540-8159.2005.40055.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  5 in total

1.  Minimally Invasive Implantation of a Micropacemaker Into the Pericardial Space.

Authors:  Yaniv Bar-Cohen; Michael J Silka; Allison C Hill; Jay D Pruetz; Ramen H Chmait; Li Zhou; Sara M Rabin; Viktoria Norekyan; Gerald E Loeb
Journal:  Circ Arrhythm Electrophysiol       Date:  2018-07

2.  Spatiotemporal effects of a controlled-release anti-inflammatory drug on the cellular dynamics of host response.

Authors:  Tram T Dang; Kaitlin M Bratlie; Said R Bogatyrev; Xiao Y Chen; Robert Langer; Daniel G Anderson
Journal:  Biomaterials       Date:  2011-03-23       Impact factor: 12.479

3.  Foreign Body Reaction to Implantable Biosensors: Effects of Tissue Trauma and Implant Size.

Authors:  Yan Wang; Santhisagar Vaddiraju; Bing Gu; Fotios Papadimitrakopoulos; Diane J Burgess
Journal:  J Diabetes Sci Technol       Date:  2015-08-25

4.  Enhanced function of immuno-isolated islets in diabetes therapy by co-encapsulation with an anti-inflammatory drug.

Authors:  Tram T Dang; Anh V Thai; Joshua Cohen; Jeremy E Slosberg; Karolina Siniakowicz; Joshua C Doloff; Minglin Ma; Jennifer Hollister-Lock; Katherine M Tang; Zhen Gu; Hao Cheng; Gordon C Weir; Robert Langer; Daniel G Anderson
Journal:  Biomaterials       Date:  2013-05-07       Impact factor: 12.479

5.  Suppression of Fibrotic Reactions of Chitosan-Alginate Microcapsules Containing Porcine Islets by Dexamethasone Surface Coating.

Authors:  Min Jung Kim; Heon-Seok Park; Ji-Won Kim; Eun-Young Lee; Marie Rhee; Young-Hye You; Gilson Khang; Chung-Gyu Park; Kun-Ho Yoon
Journal:  Endocrinol Metab (Seoul)       Date:  2021-02-24
  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.