Literature DB >> 15808157

Intraobserver repeatability of macula measurements by confocal scanning laser tomography.

Aristophanis Pallikaris1, Dimitra Skondra, Miltiadis Tsilimbaris.   

Abstract

PURPOSE: To assess the repeatability of macula measurements provided by the Heidelberg Retina Tomograph II (HRT II) in normal subjects, patients with diabetic macular edema (DME), and patients with choroidal neovascularization (CNV).
DESIGN: Prospective, observational case.
SETTING: Institutional-based study. PATIENT POPULATION: Ten normal eyes, 14 eyes with DME, and 14 eyes with CNV. OBSERVATIONAL PROCEDURE: Three scans within a single day. An additional three scans of each of the 10 normal eyes on a different day. MAIN OUTCOME MEASURES: Bland and Altman analysis for a 0.25 mm, 0.50 mm, and 0.75 mm radius circle.
RESULTS: Variability and repeatability of the measurements were not related to circle size for either group of patients. Coefficient of variability values of the normal group and a 0.50-mm radius circle were 17%, 5.7%, and 6.1% for mean surface height, signal width, and edema index, respectively. The corresponding values were 15.6%, 10%, and 10% for the DME group, and 10.5%, 7.4%, and 8.5% for the CNV group. The 95% confidence interval of the difference between single observations across days of the normal group was +/-106 microm, +/-80 microm, and +/-0.10 for mean surface height, signal width, and edema index, respectively. When three scans were combined, the corresponding values were +/-61.5 microm, +/-46.7 microm, and +/-0.06.
CONCLUSIONS: The repeatability and variability of measurements of normal subjects, patients with DME, and patients with CNV were determined. Differences between the three groups exist and are important for diagnosing a pathologic condition or measuring the effect of various treatments.

Entities:  

Mesh:

Year:  2005        PMID: 15808157     DOI: 10.1016/j.ajo.2004.11.063

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  2 in total

1.  [Measurement of retinal thickness with the Heidelberg retina tomograph for patients with macular edema and healthy individuals].

Authors:  A Ackermann; L E Pillunat
Journal:  Ophthalmologe       Date:  2012-09       Impact factor: 1.059

Review 2.  A reference standard for the measurement of macular oedema.

Authors:  K A Goatman
Journal:  Br J Ophthalmol       Date:  2006-09       Impact factor: 4.638