Literature DB >> 15799796

Cortical surface tracking using a stereoscopic operating microscope.

Hai Sun1, David W Roberts, Hany Farid, Ziji Wu, Alex Hartov, Keith D Paulsen.   

Abstract

OBJECTIVE: To measure and compensate for soft tissue deformation during image-guided neurosurgery, we have developed a novel approach to estimate the three-dimensional (3-D) topology of the cortical surface and track its motion over time.
METHODS: We use stereopsis to estimate the 3-D cortical topology during neurosurgical procedures. To facilitate this process, two charge-coupled device cameras have been attached to the binocular optics of a stereoscopic operating microscope. Before surgery, this stereo imaging system is calibrated to obtain the extrinsic and intrinsic camera parameters. During surgery, the 3-D shape of the cortical surface is automatically estimated from a stereo pair of images and registered to the preoperative image volume to provide navigational guidance. This estimation requires robust matching of features between the images, which, when combined with the camera calibration, yields the desired 3-D coordinates. After the 3-D cortical surface has been estimated from stereo pairs, its motion is tracked by comparing the current surface with its previous locations.
RESULTS: We are able to estimate the 3-D topology of the cortical surface with an average error of less than 1.2 mm. Executing on a 1.1-GHz Pentium machine, the 3-D estimation from a stereo pair of 1024 x 768 resolution images requires approximately 60 seconds of computation. By applying stereopsis over time, we are able to track the motion of the cortical surface, including the pulsatile movement of the cortical surface, gravitational sag, tissue bulge as a result of increased intracranial pressure, and the parenchymal shape changes associated with tissue resection. The results from 10 surgical patients are reported.
CONCLUSION: We have demonstrated that a stereo vision system coupled to the operating microscope can be used to efficiently estimate the dynamic topology of the cortical surface during surgery. The 3-D surface can be coregistered to the preoperative image volume. This unique intraoperative imaging technique expands the capability of the current navigational system in the operating room and increases the accuracy of anatomic correspondence with preoperative images through compensation for brain deformation.

Entities:  

Mesh:

Year:  2005        PMID: 15799796     DOI: 10.1227/01.neu.0000146263.98583.cc

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  21 in total

1.  Design and evaluation of an optically-tracked single-CCD laser range scanner.

Authors:  Thomas S Pheiffer; Amber L Simpson; Brian Lennon; Reid C Thompson; Michael I Miga
Journal:  Med Phys       Date:  2012-02       Impact factor: 4.071

2.  A fast and efficient method to compensate for brain shift for tumor resection therapies measured between preoperative and postoperative tomograms.

Authors:  Prashanth Dumpuri; Reid C Thompson; Aize Cao; Siyi Ding; Ishita Garg; Benoit M Dawant; Michael I Miga
Journal:  IEEE Trans Biomed Eng       Date:  2010-02-17       Impact factor: 4.538

3.  An atlas-based method to compensate for brain shift: preliminary results.

Authors:  Prashanth Dumpuri; Reid C Thompson; Benoit M Dawant; A Cao; Michael I Miga
Journal:  Med Image Anal       Date:  2007-03-01       Impact factor: 8.545

4.  Intraoperative brain shift compensation: accounting for dural septa.

Authors:  Ishita Chen; Aaron M Coffey; Siyi Ding; Prashanth Dumpuri; Benoit M Dawant; Reid C Thompson; Michael I Miga
Journal:  IEEE Trans Biomed Eng       Date:  2010-11-22       Impact factor: 4.538

5.  Intraoperative laser speckle contrast imaging with retrospective motion correction for quantitative assessment of cerebral blood flow.

Authors:  Lisa M Richards; Erica L Towle; Douglas J Fox; Andrew K Dunn
Journal:  Neurophotonics       Date:  2014-08-18       Impact factor: 3.593

6.  Laser range scanning for image-guided neurosurgery: investigation of image-to-physical space registrations.

Authors:  Aize Cao; R C Thompson; P Dumpuri; B M Dawant; R L Galloway; S Ding; M I Miga
Journal:  Med Phys       Date:  2008-04       Impact factor: 4.071

7.  Estimation of brain deformation for volumetric image updating in protoporphyrin IX fluorescence-guided resection.

Authors:  Pablo A Valdés; Xiaoyao Fan; Songbai Ji; Brent T Harris; Keith D Paulsen; David W Roberts
Journal:  Stereotact Funct Neurosurg       Date:  2009-11-12       Impact factor: 1.875

8.  Brain-skull contact boundary conditions in an inverse computational deformation model.

Authors:  Songbai Ji; David W Roberts; Alex Hartov; Keith D Paulsen
Journal:  Med Image Anal       Date:  2009-06-23       Impact factor: 8.545

9.  Near Real-Time Computer Assisted Surgery for Brain Shift Correction Using Biomechanical Models.

Authors:  Kay Sun; Thomas S Pheiffer; Amber L Simpson; Jared A Weis; Reid C Thompson; Michael I Miga
Journal:  IEEE J Transl Eng Health Med       Date:  2014-04-30       Impact factor: 3.316

10.  Virtual reality augmentation in skull base surgery.

Authors:  Steffen K Rosahl; Alireza Gharabaghi; Ulrich Hubbe; Ramin Shahidi; Madjid Samii
Journal:  Skull Base       Date:  2006-05
View more