Flexural strength and moduli of hypoallergenic denture base materials.

Peter Pfeiffer1, Christian Rolleke, Lamia Sherif.   

Abstract

STATEMENT OF PROBLEM: Hypoallergenic denture base materials show no residual methyl methacrylate (MMA) or significantly lower residual MMA monomer content compared to polymethyl methacrylate-based (PMMA) heat-polymerizing acrylic resin. There is insufficient knowledge of the mechanical properties of hypoallergenic denture base materials to warrant their use in place of PMMA-based acrylic resins for patients with allergic reaction to MMA.
PURPOSE: This in vitro study compared flexural strength and flexural modulus of 4 hypoallergenic denture base materials with flexural strength/modulus of a PMMA heat-polymerizing acrylic resin.
MATERIAL AND METHODS: The following denture base resins were examined: Sinomer (heat-polymerized, modified methacrylate), Polyan (thermoplastic, modified methacrylate), Promysan (thermoplastic, enterephthalate-based), Microbase (microwave-polymerized, polyurethane-based), and Paladon 65 (heat-polymerized, methacrylate, control group). Specimens of each material were tested for flexural strength and flexural modulus (MPa, n = 5) according to ISO 1567:1999. The data were analyzed with 1-way analysis of variance and the Bonferroni-Dunn multiple comparisons post hoc analysis for each test variable (alpha=.05).
RESULTS: Flexural strength of Microbase (67.2 +/- 5.3 MPa) was significantly lower than Paladon 65 (78.6 +/- 5.5 MPa, P <.0001). Flexural strength of Polyan (79.7 +/- 4.2 MPa, P =.599), Promysan (83.5 +/- 3.8 MPa, P =.412), and Sinomer (72.3 +/- 2.1 MPa, P =.015) did not differ significantly from the control group. Significantly lower flexural modulus was obtained from Sinomer (1720 +/- 30 MPa, P =.0007) compared to the PMMA control group (2050 +/- 40 MPa), whereas the flexural modulus of Promysan (2350 +/- 170 MPa, P =.0005) was significantly higher than the PMMA material. Microbase (2100 +/- 210 MPa, P =.373) and Polyan (2070 +/- 60 MPa, P =.577) exhibited flexural modulus similar to the PMMA material. The tested denture base materials fulfilled the requirements regarding flexural strength (>65 MPa). With the exception of Sinomer, the tested denture base resins passed the requirements of ISO 1567 regarding flexural modulus (>2000 MPa).
CONCLUSION: Flexural modulus of Promysan was significantly higher than the PMMA material. Microbase and Sinomer exhibited significantly lower flexural strength and flexural modulus, respectively, than PMMA. The other groups did not differ significantly from the control group.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15798688     DOI: 10.1016/j.prosdent.2005.01.011

Source DB:  PubMed          Journal:  J Prosthet Dent        ISSN: 0022-3913            Impact factor:   3.426


  9 in total

1.  Effect of Disinfectants on Mechanical Properties of Orthodontic Acrylics.

Authors:  Tobias Bensel; Jens J Bock; Anne Kebernik; Christin Arnold; Sonia Mansour; Arne F Boeckler
Journal:  Int J Biomater       Date:  2019-04-24

2.  Biomechanical Stress in Obturator Prostheses: A Systematic Review of Finite Element Studies.

Authors:  Mohammed A Mousa; Johari Yap Abdullah; Nafij B Jamayet; Mohammad Khursheed Alam; Adam Husein
Journal:  Biomed Res Int       Date:  2021-08-16       Impact factor: 3.411

3.  In Vitro Analysis of the Mechanical Properties of Hypoallergenic Denture Base Resins.

Authors:  Sebastian Hinz; Tobias Bensel; Wolfgang Bömicke; Arne F Boeckler
Journal:  Materials (Basel)       Date:  2022-05-18       Impact factor: 3.748

4.  Comparative evaluation of impact and flexural strength of four commercially available flexible denture base materials: an in vitro study.

Authors:  Pande Neelam Abhay; Shori Karishma
Journal:  J Indian Prosthodont Soc       Date:  2012-11-20

5.  Flexural properties of a light-cure and a self-cure denture base materials compared to conventional alternatives.

Authors:  Emre Mumcu; Altug Cilingir; Burc Gencel; Tonguc Sülün
Journal:  J Adv Prosthodont       Date:  2011-09-25       Impact factor: 1.904

6.  Water Sorption and Flexural Strength of Thermoplastic and Conventional Heat-Polymerized Acrylic Resins.

Authors:  Mohammad Ali Hemmati; Fariborz Vafaee; Hanif Allahbakhshi
Journal:  J Dent (Tehran)       Date:  2015-07

7.  Determination of Polyetheretherketone (PEEK) mechanical properties as a denture material.

Authors:  Saja A Muhsin; Paul V Hatton; Anthony Johnson; Nuno Sereno; Duncan J Wood
Journal:  Saudi Dent J       Date:  2019-03-13

8.  The Influence of Thermocycling on the Flexural Strength of a Polyamide Denture Base Material.

Authors:  Nick Polychronakis; Aspasia Sarafianou; Alcibiadis Zissis; Triantafillos Papadopoulos
Journal:  Acta Stomatol Croat       Date:  2017-12

9.  Effect of zinc oxide nanoparticles on the flexural strength of polymethylmethacrylate denture base resin.

Authors:  Santhanam Vikram; N Gopi Chander
Journal:  Eur Oral Res       Date:  2020-01-01
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.