Literature DB >> 15798230

Automatic detection of detached and erroneous electrodes in electrical impedance tomography.

Yednek Asfaw1, Andy Adler.   

Abstract

One unfortunate occurrence in experimental measurements with electrical impedance tomography is electrodes which become detached or poorly connected, such that the measured data cannot be used. This paper presents an automatic approach to detect such erroneous electrodes. It is based on the assumption that all valid measurements are related by the image reconstruction model, while the measurements from erroneous electrodes are unrelated. The method estimates the data at an electrode based on the measurements from all other electrodes, and compares it to the measurements. If these data match adequately, the set of electrodes does not contain an erroneous electrode. In order to detect an erroneous electrode amongst N electrodes, all sets of N-1 electrodes are tested, and the set with the best match between measurements and estimate is identified as the one which excludes the erroneous electrode. The method was tested on simulated and experimental data and showed consistent identification of erroneous electrodes with those made by experts.

Mesh:

Year:  2005        PMID: 15798230     DOI: 10.1088/0967-3334/26/2/017

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  4 in total

1.  Fast detection and data compensation for electrodes disconnection in long-term monitoring of dynamic brain electrical impedance tomography.

Authors:  Ge Zhang; Meng Dai; Lin Yang; Weichen Li; Haoting Li; Canhua Xu; Xuetao Shi; Xiuzhen Dong; Feng Fu
Journal:  Biomed Eng Online       Date:  2017-01-07       Impact factor: 2.819

2.  An on-line processing strategy for head movement interferences removal of dynamic brain electrical impedance tomography based on wavelet decomposition.

Authors:  Ge Zhang; Weichen Li; Hang Ma; Xuechao Liu; Meng Dai; Canhua Xu; Haoting Li; Xiuzhen Dong; Xingwang Sun; Feng Fu
Journal:  Biomed Eng Online       Date:  2019-05-09       Impact factor: 2.819

Review 3.  Robust imaging using electrical impedance tomography: review of current tools.

Authors:  Benoit Brazey; Yassine Haddab; Nabil Zemiti
Journal:  Proc Math Phys Eng Sci       Date:  2022-02-02       Impact factor: 2.704

4.  Real-Time Detection of Hemothorax and Monitoring its Progression in a Piglet Model by Electrical Impedance Tomography: A Feasibility Study.

Authors:  Lin Yang; Chao Zhang; Wenbo Liu; Hang Wang; Junying Xia; Benyuan Liu; Xuetao Shi; Xiuzhen Dong; Feng Fu; Meng Dai
Journal:  Biomed Res Int       Date:  2020-02-27       Impact factor: 3.411

  4 in total