Literature DB >> 15794793

A stereological analysis of fibrosis and inflammatory reaction induced by four different synthetic slings.

Marcelo Thiel1, Paulo C Rodrigues Palma, Cássio L Z Riccetto, Miriam Dambros, Nelson R Netto.   

Abstract

OBJECTIVES: To analyse quantitatively, using stereological methods, the density of the collagen fibres induced by four types of sling materials, and verify by a histopathological analysis the corresponding inflammatory reaction, as fibrosis secondary to sling implantation is considered responsible for restoring urethral support and re-establishing continence in women with stress urinary incontinence, and new synthetic materials that promote adequate fibrosis with the least intensity and duration have been proposed to substitute the aponeurotic sling.
MATERIALS AND METHODS: The study comprised 70 isogenic white Wistar rats divided into three groups: group A (30 rats) had 8 x 4 mm strips of silicone and porcine small intestine submucosa (SIS) implanted in the abdominal subcutaneous tissues; group B (30 rats) had 8 x 4 mm strips of polycaprolactone and polylactic acid copolymers and monofilament polypropylene (PLP) implanted the abdominal subcutaneous layer; while a control group of 10 rats had dissection and suturing with 5/0 Nylon in the abdominal subcutaneous layer, as used to fix the strips in the other rats. Picro-Sirius staining was used to assess collagen fibres, and haematoxylin-eosin for the histopathological study. At 7, 30 and 90 days after surgery, 10 rats from each group were killed and assessed.
RESULTS: After 7 days all the materials induced a moderate inflammatory reaction that did not differ from that in the control group. At 30 days there was no difference between the control and polycaprolactone and polylactic acid copolymers, having the least inflammatory reaction. PLP and silicone produced a moderate inflammatory reaction, while the porcine SIS induced a more intense reaction. At 90 days there was a more intense inflammatory reaction in polycaprolactone and polylactic acid copolymers than before but it was less than with PLP and silicone, which again were no different. During this period the inflammatory reaction induced by SIS was greater. The stereological analysis indicated that collagen fibres induced by polycaprolactone and polylactic acid copolymers and PLP were less dense (61% and 65%, respectively), and significantly less than with silicone (85%) and SIS (86%).
CONCLUSION: PLP was the best nonabsorbable material as it induced a less intense inflammatory reaction than the other tested materials. As porcine SIS was completely absorbed the intense fibrosis induced is useful, as it is exclusively responsible for the urethral support later after surgery.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15794793     DOI: 10.1111/j.1464-410X.2005.05411.x

Source DB:  PubMed          Journal:  BJU Int        ISSN: 1464-4096            Impact factor:   5.588


  10 in total

Review 1.  Pubovaginal sling materials and their outcomes.

Authors:  Ömer Bayrak; David Osborn; William Stuart Reynolds; Roger Roman Dmochowski
Journal:  Turk J Urol       Date:  2014-12

Review 2.  The use of synthetic sub-urethral slings in the treatment of female stress urinary incontinence.

Authors:  Andrew Feifer; Jacques Corcos
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2007-04-27

Review 3.  Polypropylene mesh and the host response.

Authors:  Hiren Patel; Donald R Ostergard; Gina Sternschuss
Journal:  Int Urogynecol J       Date:  2012-03-20       Impact factor: 2.894

4.  Can highly purified collagen coating modulate polypropylene mesh immune-inflammatory and fibroblastic reactions? Immunohistochemical analysis in a rat model.

Authors:  Fernando Goulart Fernandes Dias; Alessandro Prudente; Rodrigo Teixeira Siniscalchi; Benedicto Campos de Vidal; Cassio Luis Zanettini Riccetto
Journal:  Int Urogynecol J       Date:  2014-10-22       Impact factor: 2.894

5.  Collagen remodeling and suburethral inflammation might account for preserved anti-incontinence effects of cut polypropylene sling in rat model.

Authors:  Chi Chiung Grace Chen; Adonis Hijaz; Judith A Drazba; Margot S Damaser; Firouz Daneshgari
Journal:  Urology       Date:  2008-10-01       Impact factor: 2.649

6.  Impact of the structure of polypropylene meshes in local tissue reaction: in vivo stereological study.

Authors:  Cassio Riccetto; Ricardo Miyaoka; Rogerio de Fraga; Regina Barbosa; Miriam Dambros; Andreza Teixeira; Paulo Palma
Journal:  Int Urogynecol J Pelvic Floor Dysfunct       Date:  2008-03-20

7.  Highly purified collagen coating enhances tissue adherence and integration properties of monofilament polypropylene meshes.

Authors:  Rodrigo Teixeira Siniscalchi; Marli Melo; Paulo César Rodrigues Palma; Inácio Maria Dal Fabbro; Benedicto de Campos Vidal; Cassio Luiz Zanettini Riccetto
Journal:  Int Urogynecol J       Date:  2013-05-04       Impact factor: 2.894

Review 8.  Biomaterials for pelvic floor reconstructive surgery: how can we do better?

Authors:  Giulia Gigliobianco; Sabiniano Roman Regueros; Nadir I Osman; Julio Bissoli; Anthony J Bullock; Chris R Chapple; Sheila MacNeil
Journal:  Biomed Res Int       Date:  2015-04-21       Impact factor: 3.411

9.  Acute in vivo response to an alternative implant for urogynecology.

Authors:  Sabiniano Roman Regueros; Maarten Albersen; Stefano Manodoro; Silvia Zia; Nadir I Osman; Anthony J Bullock; Christopher R Chapple; Jan Deprest; Sheila MacNeil
Journal:  Biomed Res Int       Date:  2014-07-17       Impact factor: 3.411

10.  A novel surgical technique for a rat subcutaneous implantation of a tissue engineered scaffold.

Authors:  Reza Khorramirouz; Jason L Go; Christopher Noble; Soumen Jana; Eva Maxson; Amir Lerman; Melissa D Young
Journal:  Acta Histochem       Date:  2018-03-05       Impact factor: 2.479

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.