The sliding slit test for dynamic IMRT: a useful tool for adjustment of MLC related parameters.

I Chauvet1, A Petitfils, C Lehobey, J Y Kristner, Y Brunet, R Lembrez, G Gaboriaud, A Mazal, S Zefkili, J C Rosenwald.   

Abstract

For treatments with dynamic intensity modulated radiotherapy (IMRT), the adjustment of multileaf collimator (MLC) parameters affecting both the optimization algorithm and dose distributions is crucial. The main parameters characterizing the MLC are the transmission (T) and the dosimetric leaf separation (DLS). The aim of this study is twofold: a methodology based on the 'sliding slit' test is proposed to determine (T, DLS) combinations inducing the best conformity between calculations and measurements. Secondly, the effects of the MLC adjustment on measured dose and on optimization are presented for different configurations as the chair test and for the patient dosimetric quality control (DQC). Tests were performed with a Varian 23EX linac operated at 20 MV and equipped with a 120 leaf Millenium dynamic collimator. The treatment planning system was CadPlan/Helios (version 6.3.6). Results demonstrated that the sliding width (SW) strongly depends on the (T, DLS) combinations, and the measured dose is a linear function of the SW. Different (T, DLS) combinations induced a good agreement between calculations and measurements. The influence of the MLC calibration was found to be particularly important on the 'sliding slit' test (11.8% for a gap change of 0.8 mm) but not so much on the chair test and on the DQC. To detect small variations in leaf adjustment and to ensure consistency between calculation and actual dose delivered to patients, a daily check called IMRT MU check is proposed.

Entities:  

Mesh:

Year:  2005        PMID: 15773620     DOI: 10.1088/0031-9155/50/4/001

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  9 in total

1.  Evaluation of the Differences Between Measurements in Multiple Institutions and Calculation Modeled by Representative Beam Data in Prostate VMAT Plan.

Authors:  Hironao Goto; Hirokazu Mizuno; Yuichi Akino; Masaru Isono; Yoshihiro Tanaka; Norihisa Masai; Toshijiro Yamamoto; Masahiko Koizumi
Journal:  In Vivo       Date:  2020 May-Jun       Impact factor: 2.155

2.  Validation of a modern second-check dosimetry system using a novel verification phantom.

Authors:  Daniel G McDonald; Dustin J Jacqmin; Christopher J Mart; Nicholas C Koch; Jean L Peng; Michael S Ashenafi; Mario A Fugal; Kenneth N Vanek
Journal:  J Appl Clin Med Phys       Date:  2017-01       Impact factor: 2.102

3.  The dosimetric impact of control point spacing for sliding gap MLC fields.

Authors:  Benjamin J Zwan; Jonathan Hindmarsh; Erin Seymour; Kankean Kandasamy; Kirbie Sloan; Rajesakar David; Christopher Lee
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

4.  Relationship between dosimetric leaf gap and dose calculation errors for high definition multi-leaf collimators in radiotherapy.

Authors:  Jinkoo Kim; James S Han; An Ting Hsia; Shidong Li; Zhigang Xu; Samuel Ryu
Journal:  Phys Imaging Radiat Oncol       Date:  2018-02-20

5.  A novel methodology for the optimization of transmission and dosimetric leaf gap parameters.

Authors:  Dominic J DiCostanzo; Ahmet S Ayan
Journal:  J Appl Clin Med Phys       Date:  2022-02-28       Impact factor: 2.243

6.  Portal dosimetry correction method for validation of single isocenter VMAT plans for multiple brain metastases.

Authors:  Ronald J Lalonde; M Saiful Huq
Journal:  J Appl Clin Med Phys       Date:  2022-08-12       Impact factor: 2.243

7.  Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study.

Authors:  P Fenoglietto; M Khodri; D Nguyen; F Josserand-Pietri; N Aillères
Journal:  Radiat Oncol       Date:  2016-01-14       Impact factor: 3.481

8.  IMRT and RapidArc commissioning of a TrueBeam linear accelerator using TG-119 protocol cases.

Authors:  Ning Wen; Bo Zhao; Jinkoo Kim; Karen Chin-Snyder; Maria Bellon; Carri Glide-Hurst; Kenneth Barton; Daiquan Chen; Indrin J Chetty
Journal:  J Appl Clin Med Phys       Date:  2014-09-08       Impact factor: 2.102

9.  Characteristics of a novel treatment system for linear accelerator-based stereotactic radiosurgery.

Authors:  Ning Wen; Haisen Li; Kwang Song; Karen Chin-Snyder; Yujiao Qin; Jinkoo Kim; Maria Bellon; Misbah Gulam; Stephen Gardner; Anthony Doemer; Suneetha Devpura; James Gordon; Indrin Chetty; Farzan Siddiqui; Munther Ajlouni; Robert Pompa; Zane Hammoud; Michael Simoff; Steven Kalkanis; Benjamin Movsas; M Salim Siddiqui
Journal:  J Appl Clin Med Phys       Date:  2015-07-08       Impact factor: 2.102

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.