Thumnoon Nhujak1, Monpichar Srisa-art, Kanyarat Kalampakorn, Vasana Tolieng, Amorn Petsom.
Abstract
Micellar electrokinetic chromatography (MEKC) was developed as a method for quantitative determination of gibberellic acid (GA3) in fermentation broth and commercial products, using 25 mM disodium tetraborate as a buffer at pH 9.2 and 100 mM sodium dodecyl sulfate as a micellar phase. The baseline resolution (Rs of GA3 from other compounds in fermentation broth was achieved with Rs > 2.5. The addition of methanol or acetonitrile in the MEKC buffer did not give a better resolution. Advantages of this MEKC method include high accuracy and precision and no sample preparation except for dilution and filtration.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15769108 DOI: 10.1021/jf0484733