Literature DB >> 15768364

Distal tibiofibular syndesmosis fixation: a cadaveric, simulated fracture stabilization study comparing bioabsorbable and metallic single screw fixation.

Stephen Cox1, Debi P Mukherjee, Alan L Ogden, Raymond H Mayuex, Kalia K Sadasivan, James A Albright, William S Pietrzak.   

Abstract

Metal screws that are used for ruptured tibiofibular syndesmosis repair are often removed within 3 months of placement, suggesting the utility of bioabsorbable screws. A biomechanical study was performed to compare fixation of a simulated syndesmosis separation with a 5-mm oriented copolymer bioabsorbable (82:18 poly-L-lactic acid/poly-glycolic acid) versus a stainless steel screw. Eight pairs of cadaveric lower-leg specimens were cleaned and a pronation external rotation-type injury was created in each. The syndesmosis was fixed with a single, tricortical bioabsorbable screw in 1 ankle and a metal screw in the contralateral ankle (matched pairs). Sequential testing of the specimens showed that torsional stiffness of the fixed, relative to intact, specimens was nearly equivalent (0.730 +/- 0.260 for copolymer, 0.770 +/- 0.300 for stainless steel; P = .401). Application of 1000 cycles of axial load (90 to 900 N) resulted in a significant decrease ( P < .0001) in axial stiffness for each fixation method, but the relative decrease was equivalent for both ( P = .211). Failure torque (17.8 +/- 8.3 N.m copolymer, 21.0 +/- 11.5 N.m stainless steel; P = .238) and angle of rotation at failure (47.9 +/- 13.6 degrees copolymer, 42.0 +/- 11.5 degrees stainless steel; P = .199) were also nearly equivalent. It appears that the 5.0-mm diameter copolymer screw is biomechanically equivalent to the 5.0-mm diameter stainless steel screw for repair of syndesmosis disruption.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15768364     DOI: 10.1053/j.jfas.2005.01.010

Source DB:  PubMed          Journal:  J Foot Ankle Surg        ISSN: 1067-2516            Impact factor:   1.286


  9 in total

Review 1.  Conservative and surgical management of acute isolated syndesmotic injuries: ESSKA-AFAS consensus and guidelines.

Authors:  C Niek van Dijk; Umile Giuseppe Longo; Mattia Loppini; Pino Florio; Ludovica Maltese; Mauro Ciuffreda; Vincenzo Denaro
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2016-02-04       Impact factor: 4.342

2.  The use of TightRope fixation for ankle syndesmosis injuries: our experience.

Authors:  M Bondi; N Rossi; A Pizzoli; L Renzi Brivio
Journal:  Musculoskelet Surg       Date:  2016-08-27

3.  Rapid cooling through the glass transition transiently increases ductility of PGA/PLLA copolymers: a proposed mechanism and implications for devices.

Authors:  William S Pietrzak
Journal:  J Mater Sci Mater Med       Date:  2007-05-05       Impact factor: 3.896

Review 4.  To retain or remove the syndesmotic screw: a review of literature.

Authors:  T Schepers
Journal:  Arch Orthop Trauma Surg       Date:  2010-12-16       Impact factor: 3.067

5.  Flexible fixation of syndesmotic diastasis using the assembled bolt-tightrope system.

Authors:  Guohui Xu; Wei Chen; Qi Zhang; Juan Wang; Yanling Su; Yingze Zhang
Journal:  Scand J Trauma Resusc Emerg Med       Date:  2013-09-22       Impact factor: 2.953

6.  COMPARATIVE STUDY BETWEEN OSTEOSYNTHESIS IN CONVENTIONAL AND BIOABSORBABLE IMPLANTS IN ANKLE FRACTURES.

Authors:  Guilherme Pelosini Gaiarsa; Paulo Roberto Dos Reis; Rames Mattar; Jorge Dos Santos Silva; Túlio Diniz Fernandez
Journal:  Acta Ortop Bras       Date:  2015 Sep-Oct       Impact factor: 0.513

7.  A comparison of double single suture-button fixation, suture-button fixation, and screw fixation for ankle syndesmosis injury: A retrospective cohort study.

Authors:  Alper Kurtoglu; Alauddin Kochai; Mustafa Erkan Inanmaz; Erhan Sukur; Dogan Keskin; Mehmet Türker; Mustafa Uysal; Zafer Sen; Ismail Daldal
Journal:  Medicine (Baltimore)       Date:  2021-04-02       Impact factor: 1.889

8.  [Comparison of bioabsorbable screw and metallic screw for Maisonneuve fracture].

Authors:  Xiaodong Wen; Jun Lu; Hongmou Zhao; Yi Li; Xin Chang; Yan Zhang; Jingqi Liang; Xiaojun Liang
Journal:  Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi       Date:  2020-04-15

9.  Comparison of tendon suture fixation and cortical screw fixation for treatment of distal tibiofibular syndesmosis injury: A case-control study.

Authors:  Liuhong Song; Zhifu Liao; Zhongqiang Kuang; Shiyan Qu; Wei Zhang; Yang Yuan; Teng Fang
Journal:  Medicine (Baltimore)       Date:  2020-08-21       Impact factor: 1.817

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.