Literature DB >> 15766826

Clinical study of interference with cardiac pacemakers by a magnetic field at power line frequencies.

Alexandre Trigano1, Olivier Blandeau, Martine Souques, Jean Pierre Gernez, Isabelle Magne.   

Abstract

OBJECTIVES: This study examined the risk of interference by high magnetic flux density with permanent pacemakers.
BACKGROUND: Several forms of electromagnetic energy may interfere with the functions of implanted pacemakers. No clinical study has reported specific and relevant information pertaining to magnetic fields near power lines or electrical appliances.
METHODS: A total of 250 consecutive tests were performed in 245 recipients of permanent pacemakers during 12-lead electrocardiographic monitoring. A dedicated exposure system generated a 50-Hz frequency and maximum 100-microT flux density, while the electrical field was kept at values on the order of 0.10 V/m.
RESULTS: A switch to the asynchronous mode was recorded in three patients with devices programmed in the unipolar sensing configuration. A sustained mode switch was followed by symptomatic pacing inhibition in one patient. No effect on devices programmed in bipolar sensing was observed, except for a single interaction with a specific capture monitoring algorithm.
CONCLUSIONS: The overall incidence of interaction by a magnetic field was low in patients tested with a wide variety of conventionally programmed pacemaker models. A magnetic field pulsed at power frequency can cause a mode switch and pacing inhibition in patients with devices programmed in the unipolar sensing configuration. The risk of interference appears negligible in patients with bipolar sensing programming.

Entities:  

Mesh:

Year:  2005        PMID: 15766826     DOI: 10.1016/j.jacc.2004.10.072

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  11 in total

1.  Experimental study on malfunction of pacemakers due to exposure to different external magnetic fields.

Authors:  Maria Tiikkaja; Tommi Alanko; Harri Lindholm; Maila Hietanen; Juha Hartikainen; Lauri Toivonen
Journal:  J Interv Card Electrophysiol       Date:  2012-01-11       Impact factor: 1.900

2.  Implantable cardioverter defibrillator and 50-Hz electric and magnetic fields exposure in the workplace.

Authors:  M Souques; I Magne; J Lambrozo
Journal:  Int Arch Occup Environ Health       Date:  2010-11-11       Impact factor: 3.015

3.  Experimental and theoretical investigation of implantable cardiac pacemaker exposed to low frequency magnetic field.

Authors:  A Babouri; A Hedjeidj; L Guendouz
Journal:  J Clin Monit Comput       Date:  2009-01-06       Impact factor: 2.502

4.  Effects of external electrical and magnetic fields on pacemakers and defibrillators: from engineering principles to clinical practice.

Authors:  Roy Beinart; Saman Nazarian
Journal:  Circulation       Date:  2013-12-24       Impact factor: 29.690

Review 5.  Deep brain stimulation and electromagnetic interference.

Authors:  Shervin Rahimpour; Musa Kiyani; Sarah E Hodges; Dennis A Turner
Journal:  Clin Neurol Neurosurg       Date:  2021-02-25       Impact factor: 1.876

Review 6.  Magnetic resonance imaging in patients with cardiac pacemakers: era of "MR Conditional" designs.

Authors:  Jerold S Shinbane; Patrick M Colletti; Frank G Shellock
Journal:  J Cardiovasc Magn Reson       Date:  2011-10-27       Impact factor: 5.364

7.  Occupational exposure to electric and magnetic fields during tasks at ground or floor level at 110 kV substations in Finland.

Authors:  Leena Korpinen; Rauno Pääkkönen
Journal:  Int J Occup Saf Ergon       Date:  2016-04-14

8.  Examination of the Effect of a 50-Hz Electromagnetic Field at 500 μT on Parameters Related With the Cardiovascular System in Rats.

Authors:  Yemao Zhang; Lijuan Li; Xingfa Liu; Lijian Ding; Xiong Wu; Jin Wang; Mengying He; Huiying Hou; Guoran Ruan; Jinsheng Lai; Chen Chen
Journal:  Front Public Health       Date:  2020-04-07

9.  Testing of common electromagnetic environments for risk of interference with cardiac pacemaker function.

Authors:  Maria Tiikkaja; Aapo L Aro; Tommi Alanko; Harri Lindholm; Heli Sistonen; Juha E K Hartikainen; Lauri Toivonen; Jukka Juutilainen; Maila Hietanen
Journal:  Saf Health Work       Date:  2013-07-20

10.  Feasibility, safety, and potential demand of emergent brain magnetic resonance imaging of patients with cardiac implantable electronic devices.

Authors:  Maki Ono; Makoto Suzuki; Mitsuaki Isobe
Journal:  J Arrhythm       Date:  2017-02-09
View more