Precision of bolus thermodilution cardiac output measurements in patients with atrial fibrillation.

M Ostergaard1, L B Nilsson, J C Nilsson, J P Rasmussen, P G Berthelsen.   

Abstract

BACKGROUND: The precision of bolus thermodilution cardiac output measurements in patients with atrial fibrillation (AF) has not previously been determined. A priori we suspected that the precision would be lower in patients with AF than in patients with sinus rhythm (SR). Consequently, we also determined if the precision could be improved by injecting the thermal indicator into the right ventricle instead of the right atrium.
METHODS: Cardiac output was determined as the average result of four injections of 10 ml of iced saline. Replicate measurements were performed with thermal indicator injections into the right atrium and ventricle. The coefficients of variation and the precisions were calculated.
RESULTS: In the 25 patients with AF, mean cardiac output was 3.96 l min(-1) (range 2.4-7.4), the coefficient of variation 0.073 (95% CI +/- 0.011), and the precision 0.38 l min(-1) (95% CI +/- 0.14) with injection into the right atrium. In the 25 patients with SR, mean cardiac output was 4.73 l min(-1) (range 2.4-7.3), the coefficient of variation 0.047(95% CI +/- 0.006), and the precision 0.38 l min(-1) (95% CI +/- 0.14). In both groups, an agreement analysis demonstrated that the injection of indicator into the right ventricle resulted in a significantly higher cardiac output [AF+0.25 (95% CI +/- 0.15) l min(-1), SR+0.29 ( +/- 0.20) l min(-1)].
CONCLUSION: The coefficient of variation for cardiac output determinations is 55% higher in patients with AF. Two measurements, separated by time or intervention, must differ by 15% in AF patients and 9% in SR patients before one can be 95% confident that a real change has taken place.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15752403     DOI: 10.1111/j.1399-6576.2005.00613.x

Source DB:  PubMed          Journal:  Acta Anaesthesiol Scand        ISSN: 0001-5172            Impact factor:   2.105


  8 in total

1.  Lithium dilution cardiac output measurement in the critically ill patient: determination of precision of the technique.

Authors:  M Cecconi; D Dawson; R M Grounds; A Rhodes
Journal:  Intensive Care Med       Date:  2008-09-18       Impact factor: 17.440

2.  Feasibility and influence of hTEE monitoring on postoperative management in cardiac surgery patients.

Authors:  S Treskatsch; F Balzer; F Knebel; M Habicher; J P Braun; M Kastrup; H Grubitzsch; K-D Wernecke; C Spies; M Sander
Journal:  Int J Cardiovasc Imaging       Date:  2015-06-06       Impact factor: 2.357

3.  Central venous pressure measurements improve the accuracy of leg raising-induced change in pulse pressure to predict fluid responsiveness.

Authors:  Karim Lakhal; Stephan Ehrmann; Isabelle Runge; Dalila Benzekri-Lefèvre; Annick Legras; Pierre François Dequin; Emmanuelle Mercier; Michel Wolff; Bernard Régnier; Thierry Boulain
Journal:  Intensive Care Med       Date:  2010-01-29       Impact factor: 17.440

Review 4.  Perioperative Management of Elderly Patients with Gastrointestinal Malignancies: The Contribution of Anesthesia.

Authors:  Rudolf Mörgeli; Kathrin Scholtz; Johannes Kurth; Sascha Treskatsch; Bruno Neuner; Susanne Koch; Lutz Kaufner; Claudia Spies
Journal:  Visc Med       Date:  2017-08-11

Review 5.  Systematic review of cardiac output measurements by echocardiography vs. thermodilution: the techniques are not interchangeable.

Authors:  Mik Wetterslev; Hasse Møller-Sørensen; Rasmus Rothmann Johansen; Anders Perner
Journal:  Intensive Care Med       Date:  2016-03-01       Impact factor: 17.440

6.  Fluid challenge: tracking changes in cardiac output with blood pressure monitoring (invasive or non-invasive).

Authors:  Karim Lakhal; Stephan Ehrmann; Dominique Perrotin; Michel Wolff; Thierry Boulain
Journal:  Intensive Care Med       Date:  2013-09-06       Impact factor: 17.440

7.  Respiratory pulse pressure variation fails to predict fluid responsiveness in acute respiratory distress syndrome.

Authors:  Karim Lakhal; Stephan Ehrmann; Dalila Benzekri-Lefèvre; Isabelle Runge; Annick Legras; Pierre-François Dequin; Emmanuelle Mercier; Michel Wolff; Bernard Régnier; Thierry Boulain
Journal:  Crit Care       Date:  2011-03-07       Impact factor: 9.097

8.  Time to correct the flow of corrected flow time.

Authors:  Igor Barjaktarevic; Alan Chiem; Maxime Cannesson
Journal:  Crit Ultrasound J       Date:  2017-10-04
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.