Beat-to-beat effects of intraaortic balloon pump timing on left ventricular performance in patients with low ejection fraction.

Jan J Schreuder1, Francesco Maisano, Andrea Donelli, Jos R C Jansen, Pat Hanlon, Jan Bovelander, Ottavio Alfieri.   

Abstract

BACKGROUND: Intraaortic balloon counterpulsation (IABP) timing errors during arrhythmia may result in afterload increases which may negatively influence left ventricular (LV) ejection and LV mechanical dyssynchrony. The aim of our study was to determine beat-to-beat effects of properly timed IABP, premature IAB inflation, and late IAB deflation on LV performance and LV mechanical dyssynchrony in heart failure patients undergoing cardiac surgery.
METHODS: In 15 patients, LV pressure-volume relations and LV dyssynchrony were measured by conductance volume catheter. Properly timed IABP was evaluated at a 1:1 assist ratio within a 10 seconds time-span. Premature IAB inflation and late IAB deflation were evaluated at a 1:4 assist ratio.
RESULTS: Properly timed 1:1 IABP acutely decreased LV end-systolic volume by 6.1% (p < 0.0001) and LV end-systolic pressure by 17.5% (p < 0.0001) due to decreased aortic impedance. Stroke volume (SV) increased by 14% (p < 0.0001), which correlated markedly with a decrease of LV mechanical dyssynchrony (p < 0.0001). The largest SV increases occurred in patients with lowest contractile state. Premature IAB inflation decreased SV by 20% (p < 0.0001) due to abrupt increase of LV afterload during late ejection. Late IAB deflation increased SV and stroke work by 18% (p < 0.0001) and 16% (p < 0.01) respectively, due to increased afterload during early ejection and decreased afterload during late ejection.
CONCLUSIONS: Left ventricular performance during IABP is causally related to changes in LV afterload, and the timing of these changes in relation to contraction or relaxation phases, to LV mechanical dyssynchrony and to contractile state.

Entities:  

Mesh:

Year:  2005        PMID: 15734398     DOI: 10.1016/j.athoracsur.2004.07.073

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  14 in total

1.  Percutaneous Mechanical Circulatory Support for Cardiogenic Shock.

Authors:  Kevin J Morine; Navin K Kapur
Journal:  Curr Treat Options Cardiovasc Med       Date:  2016-01

2.  Effects of intra-aortic balloon pump counterpulsation on left ventricular mechanoenergetics in a porcine model of acute ischemic heart failure.

Authors:  Konstantinos Malliaras; Efstratios Charitos; Nikolaos Diakos; Iraklis Pozios; Apostolos Papalois; John Terrovitis; John Nanas
Journal:  J Cardiovasc Transl Res       Date:  2014-11-07       Impact factor: 4.132

Review 3.  Mechanical Left Ventricular Unloading to Reduce Infarct Size During Acute Myocardial Infarction: Insight from Preclinical and Clinical Studies.

Authors:  Navin K Kapur; Lara Reyelt; Lija Swain; Michele Esposito; Xiaoying Qiao; Shiva Annamalai; Bart Meyns; Richard Smalling
Journal:  J Cardiovasc Transl Res       Date:  2019-04-23       Impact factor: 4.132

Review 4.  Aortic counterpulsation: C-pulse and other devices for cardiac support.

Authors:  Pallavi Solanki
Journal:  J Cardiovasc Transl Res       Date:  2014-02-20       Impact factor: 4.132

5.  Cardiogenic shock with stunned myocardium during triple-H therapy treated with intra-aortic balloon pump counterpulsation.

Authors:  Fabio Silvio Taccone; Boris Lubicz; Michael Piagnerelli; Marc Van Nuffelen; Jean-Louis Vincent; Daniel De Backer
Journal:  Neurocrit Care       Date:  2008-09-23       Impact factor: 3.210

6.  Closing the loop: modelling of heart failure progression from health to end-stage using a meta-analysis of left ventricular pressure-volume loops.

Authors:  David R Warriner; Alistair G Brown; Susheel Varma; Paul J Sheridan; Patricia Lawford; David R Hose; Abdallah Al-Mohammad; Yubing Shi
Journal:  PLoS One       Date:  2014-12-05       Impact factor: 3.240

Review 7.  IABP: history-evolution-pathophysiology-indications: what we need to know.

Authors:  H Parissis; V Graham; S Lampridis; M Lau; G Hooks; P C Mhandu
Journal:  J Cardiothorac Surg       Date:  2016-08-04       Impact factor: 1.637

Review 8.  A Rising Hope of an Artificial Heart: Left Ventricular Assisted Device - Outcome, Convenience, and Quality of Life.

Authors:  Suyeewin Thiha; Abdul Rehman Z Zaidi; Chris A Robert; Mohammed K Abbas; Bilal Haider Malik
Journal:  Cureus       Date:  2019-09-10

9.  Discerning aortic waves during intra-aortic balloon pumping and their relation to benefits of counterpulsation in humans.

Authors:  Christina Kolyva; George M Pantalos; Guruprasad A Giridharan; John R Pepper; Ashraf W Khir
Journal:  J Appl Physiol (1985)       Date:  2009-09-03

Review 10.  The role of acute circulatory support in ST-segment elevation myocardial infarction complicated by cardiogenic shock.

Authors:  Michele Esposito; Yousef Bader; Robert Pedicini; Catalina Breton; Andrew Mullin; Navin K Kapur
Journal:  Indian Heart J       Date:  2017-05-22
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.