| Literature DB >> 15723401 |
C Stehning1, P Börnert, K Nehrke, O Dössel.
Abstract
A shortcoming of today's coronary magnetic resonance angiography (MRA) is its low total scan efficiency (<5%), as only small well-defined fractions of the respiratory (50%) and cardiac (10%) cycle are used for data acquisition. These precautions are necessary to prevent blurring and artifacts related to respiratory and cardiac motion. Hence, scan times range from 4 to 9 min, which may not be tolerated by patients. To overcome this drawback, an ECG-triggered, navigator-gated free breathing radial 3D balanced FFE sequence with intra-RR motion correction is investigated in this study. Scan efficiency is increased by using a long cardiac acquisition window during the RR interval. This allows the acquisition of a number of independent k-space segments during each cardiac cycle. The intersegment motion is corrected using a self-guided epicardial fat tracking procedure in a postprocessing step. Finally, combining the motion-corrected segments forms a high-resolution image. Experiments on healthy volunteers are presented to show the basic feasibility of this approach. (c) 2005 Wiley-Liss, Inc.Entities:
Mesh:
Year: 2005 PMID: 15723401 DOI: 10.1002/mrm.20397
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668