Soft-copy reading of digital chest radiographs: effect of ambient light and automatic optimization of monitor luminance.

Martin Uffmann1, Mathias Prokop, Walter Kupper, Thomas Mang, Volker Fiedler, Cornelia Schaefer-Prokop.   

Abstract

OBJECTIVE: We sought to evaluate whether soft-copy reading of simulated pulmonary chest lesions is influenced by ambient light and automatic optimization of cathode ray tube (CRT) monitor luminance.
MATERIALS AND METHODS: Four types of simulated lesions (nodules, lines, micronodules, and patchy opacities) were superimposed over an anthropomorphic chest phantom. Lesion detection with soft-copy reading was assessed using a high-contrast grayscale 2K CRT monitor under the following conditions: (1) subdued lighting (<50 lux); (2) normal lighting conditions (450 lux) without, and (3) with a sensitivity modulation to automatically adjust the CRT luminance to the increased amount of ambient light. Reading data were analyzed according to receiver operating curve. Significance of differences was tested using an analysis of variance for repeated measures.
RESULTS: Ambient room light of 450 lux did not significantly influence the detection of nodules and patchy opacities. However, bright ambient light significantly decreased detection of micronodules (0.60 vs. 0.74) and lines (0.52 vs. 0.66) relative to subdued lighting conditions. Automatic luminance adjustment could compensate the effect of ambient light for the micronodules (0.77) but not for the lines (0.53).
CONCLUSION: Bright ambient light significantly decreases detection of small low-contrast structures. This may be partially but not completely compensated by an automatic luminance adaptation.

Entities:  

Mesh:

Year:  2005        PMID: 15714093     DOI: 10.1097/01.rli.0000153658.15272.91

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  4 in total

1.  Visual assessment method of angular performance in medical liquid-crystal displays by use of the ANG test pattern: effect of ambient illuminance and effectiveness of modified scoring.

Authors:  Yoichiro Ikushima; Junji Morishita; Hiroshi Akamine; Yasuhiko Nakamura; Noriyuki Hashimoto
Journal:  Radiol Phys Technol       Date:  2013-08-10

2.  Breast screen new South wales generally demonstrates good radiologic viewing conditions.

Authors:  BaoLin Pauline Soh; Warwick Lee; Jennifer L Diffey; Mark F McEntee; Peter L Kench; Warren M Reed; Patrick C Brennan
Journal:  J Digit Imaging       Date:  2013-08       Impact factor: 4.056

3.  Letter to the editor re: the iPad tablet computer for mobile on-call radiology diagnosis? Auditing discrepancy in CT and MRI reporting.

Authors:  Joon Bum Park; Bo Seung Kang
Journal:  J Digit Imaging       Date:  2013-04       Impact factor: 4.056

4.  Human contrast-detail performance with declining contrast.

Authors:  Alisa Walz-Flannigan; Ben Babcock; George C Kagadis; Jihong Wang; Steve G Langer
Journal:  Med Phys       Date:  2012-09       Impact factor: 4.071

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.