Tentative identification of hemoglobin variants in the Bio-Rad VARIANT II Hb A1C method.

Trefor N Higgins1, Bernice Ridley.   

Abstract

OBJECTIVE: To evaluate if peak characteristics, the retention time, and percentage of total hemoglobin of an unknown peak on the Bio-Rad VARIANT II Hb A1c method may be used to establish a tentative identification of hemoglobin variants.
METHOD: The peak characteristics, retention time, and percentage of total hemoglobin obtained on abnormal peaks found on the Bio-Rad VARIANT II Hb A1c method were tabulated and evaluated against the identification of the hemoglobin variant established by the Bio-Rad beta thalassemia HPLC method and hemoglobin electrophoresis at both acid and alkaline pH.
RESULTS: Some hemoglobin variants show specific peak characteristics, retention times, and percentage of total hemoglobin on the Bio-Rad VARIANT II Hb A1c method that allows for tentative identification of the hemoglobin variant. The retention times and percentage of hemoglobin variant for the common hemoglobin variants E, D, S, and C obtained on the Bio-Rad VARIANT II Hb A1c method are tabulated below.
CONCLUSION: Peak characteristics, retention times, and percentage of hemoglobin variant may be used to establish the presence of a hemoglobin variant and to provide a tentative identification of some hemoglobin variants on the Bio-Rad VARIANT II Hb A1c method.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15708550     DOI: 10.1016/j.clinbiochem.2004.11.004

Source DB:  PubMed          Journal:  Clin Biochem        ISSN: 0009-9120            Impact factor:   3.281


  4 in total

1.  Sickle cell trait and development of microvascular complications in diabetes mellitus.

Authors:  Anthony J Bleyer; Sri V Reddy; Leon Sujata; Gregory B Russell; Damilola Akinnifesi; Anthony J Bleyer; Donald Hire; Zak Shihabi; Mary Ann Knovich; Pirouz Daeihagh; Jorge Calles; Barry I Freedman
Journal:  Clin J Am Soc Nephrol       Date:  2010-03-18       Impact factor: 8.237

2.  Lack of precision in HbA(1c) values on variant II in cases of hemoglobin Q India.

Authors:  P K Nigam; S Sharma; R Sareen; V Paul; A Lal
Journal:  Indian J Clin Biochem       Date:  2006-09

3.  Cation Exchange High Performance Liquid Chromatography for Diagnosis of Haemoglobinopathies.

Authors:  P K Gupta; H Kumar; S Kumar; M Jaiprakash
Journal:  Med J Armed Forces India       Date:  2011-07-21

4.  Raised HIF1α during normoxia in high altitude pulmonary edema susceptible non-mountaineers.

Authors:  Poonam Soree; Rajinder K Gupta; Krishan Singh; Koundinya Desiraju; Anurag Agrawal; Praveen Vats; Abhishek Bharadwaj; T P Baburaj; Pooja Chaudhary; Vijay K Singh; Saroj Verma; Amir Chand Bajaj; Shashi Bala Singh
Journal:  Sci Rep       Date:  2016-05-23       Impact factor: 4.379