Evaluation of particle size distribution of albuterol sulfate hydrofluoroalkane administered via metered-dose inhaler with and without valved holding chambers.

Courtney Crim1, Mark Holmes, Benjamin Lee, Robb Cavanaugh, William Lincourt.   

Abstract

BACKGROUND: Administration of inhaled medications via metered-dose inhaler (MDI) to pediatric patients younger than 4 years usually requires use of a holding chamber or spacer with an attached face mask.
OBJECTIVE: To determine the particle size distribution and overall dose of albuterol from the albuterol sulfate hydrofluoroalkane delivered in conjunction with 2 US-marketed valved holding chambers (VHCs) compared with the dose delivered via MDIs without VHCs.
METHODS: Cascade impaction methods with high-performance liquid chromatography were used to evaluate the fine particle mass (FPM) of albuterol administered without and with the use of 2 commercially available VHCs.
RESULTS: Particle size distributions for the 2 VHCs and the control were similar. The mean FPM values for the 2 VHCs and the control were 32, 28, and 30 microg, respectively. Statistical comparison of the FPM shows a similar profile when differences from the albuterol hydrofluoroalkane without a spacer were evaluated.
CONCLUSIONS: In vitro results obtained under these test conditions demonstrate that all the FPM values for the VHCs tested were within 15% of the control range, a difference that is unlikely to be clinically meaningful. These results do not warrant a change in the recommended dose of albuterol hydrofluoroalkane administered when using the VHCs tested. The use of an MDI in conjunction with a VHC provides a reasonable therapeutic approach for administration of albuterol hydrofluoroalkane to young children and other patients who have difficulty administering the MDI alone.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15702821     DOI: 10.1016/S1081-1206(10)61290-2

Source DB:  PubMed          Journal:  Ann Allergy Asthma Immunol        ISSN: 1081-1206            Impact factor:   6.347


  6 in total

1.  VACuum INtubation (VACcIN) box restricts the exhaled air dispersion generated by simulated cough: description and simulation-based tests of an innovative aerosolization protective prototype.

Authors:  Rami Issa; Robert Urbanowicz; Philippe Richebé; Julie Blain; Alexandre Ferreira Benevides; Issam Tanoubi
Journal:  J Clin Monit Comput       Date:  2021-09-25       Impact factor: 1.977

2.  Switching from CFC to HFA Inhalers: What NPs and Their Patients Need to Know.

Authors:  Barbara Velsor-Friedrich; Lisa Kinsella Militello; Kelly K Zinn; Darla K DeWolff
Journal:  Am J Nurse Pract       Date:  2009-10

Review 3.  Spacer devices for inhaled therapy: why use them, and how?

Authors:  Walter Vincken; Mark L Levy; Jane Scullion; Omar S Usmani; P N Richard Dekhuijzen; Chris J Corrigan
Journal:  ERJ Open Res       Date:  2018-06-18

4.  Orally Inhaled Drug Particle Transport in Computerized Models of Laryngotracheal Stenosis.

Authors:  Dennis Onyeka Frank-Ito; Seth Morris Cohen
Journal:  Otolaryngol Head Neck Surg       Date:  2020-10-13       Impact factor: 3.497

Review 5.  A review of the in vitro and in vivo valved holding chamber (VHC) literature with a focus on the AeroChamber Plus Flow-Vu Anti-static VHC.

Authors:  Sanjeeva Dissanayake; Jason Suggett
Journal:  Ther Adv Respir Dis       Date:  2018 Jan-Dec       Impact factor: 4.031

6.  Water Uptake by Evaporating pMDI Aerosol Prior to Inhalation Affects Both Regional and Total Deposition in the Respiratory System.

Authors:  Victoria Legh-Land; Allen E Haddrell; David Lewis; Darragh Murnane; Jonathan P Reid
Journal:  Pharmaceutics       Date:  2021-06-24       Impact factor: 6.321

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.