Peripheral dose measurement with a MOSFET detector.

Martin J Butson1, Tsang Cheung, Peter K N Yu.   

Abstract

The accuracy of a MOSFET dosimetry system with respect to peripheral therapeutic doses from high-energy X-rays has been evaluated. The results have been compared with ionisation chamber measurements in the same peripheral regions of the beam. For 6 MV and 18 MV X-ray beams, the MOSFET system in the high-sensitivity mode produces reproducibility of dose measurement with relative standard deviations within 1% of the maximal dose in the beam, if the measurement is made upto 15 cm away from the beam edge. The results have shown that the MOSFET device can adequately measure peripheral doses, which would be beneficial for in vivo dose assessments in radiotherapy.

Entities:  

Mesh:

Year:  2005        PMID: 15701420     DOI: 10.1016/j.apradiso.2004.09.001

Source DB:  PubMed          Journal:  Appl Radiat Isot        ISSN: 0969-8043            Impact factor:   1.513


  3 in total

1.  Peripheral dose measurement in high-energy photon radiotherapy with the implementation of MOSFET.

Authors:  Vassiliki Vlachopoulou; Georgia Malatara; Harry Delis; Kiki Theodorou; Dimitrios Kardamakis; George Panayiotakis
Journal:  World J Radiol       Date:  2010-11-28

2.  Performance of KCl:Eu2+ storage phosphor dosimeters for low-dose measurements.

Authors:  H Harold Li; Zhiyan Xiao; Rachael Hansel; Nels Knutson; Deshan Yang
Journal:  Phys Med Biol       Date:  2013-06-05       Impact factor: 3.609

3.  Peripheral doses in patients undergoing Cyberknife treatment for intracranial lesions. A single centre experience.

Authors:  Vassiliki Vlachopoulou; Christos Antypas; Harry Delis; Argyrios Tzouras; Nikolaos Salvaras; Dimitrios Kardamakis; George Panayiotakis
Journal:  Radiat Oncol       Date:  2011-11-14       Impact factor: 3.481

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.