Literature DB >> 15689806

Refractive outcome of silicone oil removal and intraocular lens implantation using laser interferometry.

Hooshang Faghihi Habibabadi1, Hassan Hashemi, Kamran Hodjat Jalali, Ali Amini, Mohammad Riazi Esfahani.   

Abstract

PURPOSE: To evaluate the refractive outcome of silicone oil removal and intraocular lens (IOL) implantation using laser interferometry.
METHODS: Thirteen silicone oil-filled eyes of 12 patients were included in the study. IOL power calculation was performed using laser interferometry (IOLMaster V1.1; Carl Zeiss, Jena, Germany). All of these eyes underwent silicone oil removal and cataract extraction with IOL implantation. Post-operative refraction was evaluated.
RESULTS: The mean deviation of the final post-operative refraction (spherical equivalent) was -0.30+/-0.91 D (range, -1.87 to +1.3) at 12 weeks. The mean axial length of the eyes was 22.99+/-0.84 mm (range, 22.07-25.24 mm). No major complications occurred intra- or post-operatively.
CONCLUSION: Laser interferometry appears to be a feasible and satisfactorily accurate method to calculate IOL power in some silicone oil-filled eyes. Further studies comparing this technique to others are warranted.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15689806     DOI: 10.1097/00006982-200502000-00008

Source DB:  PubMed          Journal:  Retina        ISSN: 0275-004X            Impact factor:   4.256


  4 in total

1.  Accuracy and reliability of IOL master and A-scan immersion biometry in silicone oil-filled eyes.

Authors:  P Kunavisarut; P Poopattanakul; C Intarated; K Pathanapitoon
Journal:  Eye (Lond)       Date:  2012-08-10       Impact factor: 3.775

2.  Effect of anterior chamber depth on the choice of intraocular lens calculation formula in patients with normal axial length.

Authors:  Mohammad Miraftab; Hassan Hashemi; Akbar Fotouhi; Mehdi Khabazkhoob; Farhad Rezvan; Soheila Asgari
Journal:  Middle East Afr J Ophthalmol       Date:  2014 Oct-Dec

3.  Visual Outcomes and Refractive Status after Combined Silicone Oil Removal/Cataract Surgery with Intraocular Lens Implantation.

Authors:  Hussain F Al-Habboubi; Waseem Al-Zamil; Ali A Al-Habboubi; Rajiv Khandekar
Journal:  J Ophthalmic Vis Res       Date:  2018 Jan-Mar

4.  Refractive outcomes of a single-step and a two-step approach for silicone oil removal and cataract surgery.

Authors:  V G Madanagopalan; Pradeep Susvar; M Arthi
Journal:  Indian J Ophthalmol       Date:  2019-05       Impact factor: 1.848

  4 in total