Abstract

One of the main limitations of vertebroplasty is the excessive pressure required to inject a sufficient amount of cement into a vertebral body. Based on previous work that shows that approximately 95% of the injection pressure is required to deliver the cement through the cannula, we proposed a new cannula design with a larger internal diameter in the proximal section. The objective of this study is to determine whether the new cannula geometry significantly reduces the delivery pressure and eases cement injection during vertebroplasty. Two different methods were employed to examine the delivery pressure in a conventional and two redesigned cannulae: (1) analytical model: Hagen-Poisseuille's flow through a tube was used to predict the pressure drop in the cannulae; (2) experiment: first a Newtonian silicone oil and then an acrylic bone cement was injected through the cannulae at a constant rate of 4 cc/min, and the delivery pressure was recorded. Both the experimental and analytical findings confirmed that the redesigned cannula reduces the delivery pressure significantly. Specifically, when the internal diameter of the proximal section was increased by a factor of two, which is clinically feasible, the delivery pressure dropped by about 63%. The redesigned cannula appears to have the potential to improve vertebroplasty. The key benefits are that (1) it eases cement injection, (2) it can be easily integrated into the existing procedure, and (3) it is cost-effective.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15672244      PMCID: PMC3454663          DOI: 10.1007/s00586-004-0822-1

Source DB:  PubMed          Journal:  Eur Spine J        ISSN: 0940-6719            Impact factor:   3.134


  24 in total

Review 1.  Percutaneous vertebroplasty: a developing standard of care for vertebral compression fractures.

Authors:  J M Mathis; J D Barr; S M Belkoff; M S Barr; M E Jensen; H Deramond
Journal:  AJNR Am J Neuroradiol       Date:  2001-02       Impact factor: 3.825

2.  Percutaneous vertebroplasty: a special syringe for cement injection.

Authors:  I Al-Assir; A Perez-Higueras; J Florensa; A Muñoz; E Cuesta
Journal:  AJNR Am J Neuroradiol       Date:  2000-01       Impact factor: 3.825

3.  The effect of the monomer-to-powder ratio on the material properties of acrylic bone cement.

Authors:  Stephen M Belkoff; Janis C Sanders; Louis E Jasper
Journal:  J Biomed Mater Res       Date:  2002

4.  Influence of cement viscosity on cement interdigitation and venous fat content under in vivo conditions: a bilateral study of 13 sheep.

Authors:  Steffen Breusch; Christian Heisel; Jens Müller; Tanja Borchers; Hans Mau
Journal:  Acta Orthop Scand       Date:  2002-08

5.  Injection biomechanics of bone cements used in vertebroplasty.

Authors:  G Baroud; M Bohner; P Heini; T Steffen
Journal:  Biomed Mater Eng       Date:  2004       Impact factor: 1.300

6.  Effect of vertebral shell on injection pressure and intravertebral pressure in vertebroplasty.

Authors:  Gamal Baroud; Christianne Vant; Demetri Giannitsios; Marc Bohner; Thomas Steffen
Journal:  Spine (Phila Pa 1976)       Date:  2005-01-01       Impact factor: 3.468

7.  Percutaneous polymethylmethacrylate vertebroplasty in the treatment of osteoporotic vertebral body compression fractures: technical aspects.

Authors:  M E Jensen; A J Evans; J M Mathis; D F Kallmes; H J Cloft; J E Dion
Journal:  AJNR Am J Neuroradiol       Date:  1997 Nov-Dec       Impact factor: 3.825

Review 8.  Percutaneous vertebroplasty: state of the art.

Authors:  A Cotten; N Boutry; B Cortet; R Assaker; X Demondion; D Leblond; P Chastanet; B Duquesnoy; H Deramond
Journal:  Radiographics       Date:  1998 Mar-Apr       Impact factor: 5.333

9.  Percutaneous vertebroplasty and kyphoplasty for painful vertebral body fractures in cancer patients.

Authors:  Daryl R Fourney; Donald F Schomer; Remi Nader; Jennifer Chlan-Fourney; Dima Suki; Kamran Ahrar; Laurence D Rhines; Ziya L Gokaslan
Journal:  J Neurosurg       Date:  2003-01       Impact factor: 5.115

10.  Extending polymerization time of polymethylmethacrylate cement in percutaneous vertebroplasty with ice bath cooling.

Authors:  Ram Chavali; Rick Resijek; Steven K Knight; In Sup Choi
Journal:  AJNR Am J Neuroradiol       Date:  2003-03       Impact factor: 3.825

View more
  4 in total

1.  [Cement distribution in vertebroplasty pedicle screws with different designs].

Authors:  K Kafchitsas; F Geiger; M Rauschmann; S Schmidt
Journal:  Orthopade       Date:  2010-07       Impact factor: 1.087

2.  Self-setting calcium orthophosphate formulations.

Authors:  Sergey V Dorozhkin
Journal:  J Funct Biomater       Date:  2013-11-12

3.  Comparative clinical results of vertebroplasty using jamshidi® needle and bone void filler for acute vertebral compression fractures.

Authors:  Se-Il Jeon; Il-Seung Choe; Young Sub Kwon; Dae-Hee Seo; Kyu Chang Lee; Sung-Choon Park
Journal:  Korean J Spine       Date:  2012-09-30

4.  Revision strategies for complications and failure of vertebroplasties.

Authors:  Shih-Chieh Yang; Wen-Jer Chen; Shang-Won Yu; Yuan-Kun Tu; Yu-Hsien Kao; Kao-Chi Chung
Journal:  Eur Spine J       Date:  2008-04-30       Impact factor: 3.134

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.