Effect of number of projections on image quality of local CT.

A N van Daatselaar1, P F van der Stelt, J Weenen.   

Abstract

OBJECTIVES: To determine the image quality of the local CT imaging procedure, in terms of resolution, contrast, and noise as a function of the number of projections.
METHODS: The contrast and resolution of the images was determined with a phantom object consisting of three rods of different materials, as well as a phantom human head embedded in soft tissue equivalent material. In addition, slices reconstructed from computed sinograms were used for comparison.
RESULTS: Sharpness, contrast and noise were determined as a function of the number of projections. The number of projections was found to affect the contrast and the noise most, and had much less influence on resolution.
CONCLUSIONS: Judging from the images of the phantom head and the numerical data, it seems that the minimum number of projections needed to obtain images of useful quality in the geometry used is about 33. Improved image quality (at any number of projections) can best be achieved through noise suppression.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15665229     DOI: 10.1259/dmfr/23496562

Source DB:  PubMed          Journal:  Dentomaxillofac Radiol        ISSN: 0250-832X            Impact factor:   2.419


  12 in total

1.  Volumetry of human molars with flat panel-based volume CT in vitro.

Authors:  Christian Hannig; Eva Krieger; Christian Dullin; Hans-Albert Merten; Thomas Attin; Eckhardt Grabbe; Gabert Heidrich
Journal:  Clin Oral Investig       Date:  2006-05-20       Impact factor: 3.573

2.  Accuracy of linear intraoral measurements using cone beam CT and multidetector CT: a tale of two CTs.

Authors:  R Patcas; G Markic; L Müller; O Ullrich; T Peltomäki; C J Kellenberger; C A Karlo
Journal:  Dentomaxillofac Radiol       Date:  2012-05-03       Impact factor: 2.419

3.  Effect of exposure time on the accuracy and reliability of cone beam computed tomography in the assessment of dental implant site dimensions in dry skulls.

Authors:  Asmáa A Al-Ekrish
Journal:  Saudi Dent J       Date:  2012-06-01

4.  A comparative study of the accuracy and reliability of multidetector computed tomography and cone beam computed tomography in the assessment of dental implant site dimensions.

Authors:  A A Al-Ekrish; M Ekram
Journal:  Dentomaxillofac Radiol       Date:  2011-02       Impact factor: 2.419

5.  Influence of scan setting selections on root canal visibility with cone beam CT.

Authors:  B A Hassan; J Payam; B Juyanda; P van der Stelt; P R Wesselink
Journal:  Dentomaxillofac Radiol       Date:  2012-12       Impact factor: 2.419

6.  Influence of cone beam CT scanning parameters on grey value measurements at an implant site.

Authors:  A Parsa; N Ibrahim; B Hassan; A Motroni; P van der Stelt; D Wismeijer
Journal:  Dentomaxillofac Radiol       Date:  2012-08-29       Impact factor: 2.419

7.  Noise, sampling, and the number of projections in cone-beam CT with a flat-panel detector.

Authors:  Z Zhao; G J Gang; J H Siewerdsen
Journal:  Med Phys       Date:  2014-06       Impact factor: 4.071

8.  Detection of vertical root fractures in intact and endodontically treated premolar teeth by designing a probabilistic neural network: an ex vivo study.

Authors:  Masume Johari; Farzad Esmaeili; Alireza Andalib; Shabnam Garjani; Hamidreza Saberkari
Journal:  Dentomaxillofac Radiol       Date:  2016-10-27       Impact factor: 2.419

9.  Current status of dental caries diagnosis using cone beam computed tomography.

Authors:  Young-Seok Park; Jin-Soo Ahn; Ho-Beom Kwon; Seung-Pyo Lee
Journal:  Imaging Sci Dent       Date:  2011-06-23

10.  A Novel Thresholding Based Algorithm for Detection of Vertical Root Fracture in Nonendodontically Treated Premolar Teeth.

Authors:  Masume Johari; Farzad Esmaeili; Alireza Andalib; Shabnam Garjani; Hamidreza Saberkari
Journal:  J Med Signals Sens       Date:  2016 Apr-Jun
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.