Intra-aortic balloon pumping: does posture matter?

Ashraf W Khir1, Susana Price, Colin Hale, David A Young, Kim H Parker, John R Pepper.   

Abstract

BACKGROUND: The effect of the posture of semirecumbent patients on the hemodynamics and performance of intra-aortic balloon (IAB) were studied in vivo and in vitro.
METHODS: An IAB was inserted into a glass tube filled with saline, fully inflated and deflated using an intra-aortic balloon pump (IABP). Three successive cycles were filmed at 125 frames/s with the tube positioned at various angles between 0 degrees (horizontal) and 90 degrees (vertical). Pressure and flow were measured distal to both ends of the balloon. In parallel, coronary left anterior descending (LAD) flow velocity and aortic pressure were recorded in 6 patients using IABP, postcardiac surgery in the intensive care unit. Recordings were made when the patient was lying horizontally (recumbent) and when the patient's torso was inclined at 30 degrees to the horizontal (semirecumbent).
RESULTS: With the tube horizontal, the inflation was effectively uniform along the length of the balloon. At all other angles, the higher end of the balloon inflated first, and mean pressure and flow measured distal to the higher end of the balloon were less than those measured at 0 degrees . Mean aortic pressure and LAD flow decreased by 10 +/- 2% (P = 0.001) and 15 +/- 3% (P = 0.001), respectively, when the patient was semirecumbent compared to when the patient was recumbent.
CONCLUSION: The decreased mean aortic pressure and LAD flow velocity suggests that unless patients using IABP are required to be semirecumbent, it may be best to position them horizontally to gain the full benefits of balloon counter pulsation to the coronary circulation.

Entities:  

Mesh:

Year:  2005        PMID: 15644081     DOI: 10.1111/j.1525-1594.2004.29013.x

Source DB:  PubMed          Journal:  Artif Organs        ISSN: 0160-564X            Impact factor:   3.094


  3 in total

Review 1.  Evidence of clinical efficacy of counterpulsation therapy methods.

Authors:  M Capoccia; C T Bowles; J R Pepper; N R Banner; A R Simon
Journal:  Heart Fail Rev       Date:  2015-05       Impact factor: 4.214

2.  Newly Shaped Intra-Aortic Balloons Improve the Performance of Counterpulsation at the Semirecumbent Position: An In Vitro Study.

Authors:  Christina Kolyva; John R Pepper; Ashraf W Khir
Journal:  Artif Organs       Date:  2016-08       Impact factor: 3.094

3.  Measurements of Intra-Aortic Balloon Wall Movement During Inflation and Deflation: Effects of Angulation.

Authors:  Gianpaolo Bruti; Christina Kolyva; John R Pepper; Ashraf W Khir
Journal:  Artif Organs       Date:  2015-05-08       Impact factor: 3.094

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.