| Literature DB >> 15626616 |
Galina Kurlyandskaya1, Vladimir Levit.
Abstract
A GMI-biosensor prototype was designed and fabricated using an amorphous ribbon. As a first step, a GMI response was measured with a model liquid media: suspension containing magnetisable Dynabeads M-450. These beads can be used as biomolecular labels when coated with a specific antibody. The GMI change caused by their presence was measured with a biosensor prototype designed with a measuring cell containing an amorphous CoFeMoSiB-ribbon element. GMI was measured at a range of current frequencies from 0.3 to 10 MHz and at intensities of either Irms=3 or 4.5 mA. Commercial Dynabeads M-450 were supplied as a suspension in a phosphate buffered saline. A maximum difference of 25% in the GMI ratio measured with and without Dynabeads was obtained at a frequency of 3.5 MHz and Irms=3 mA. Some potential applications for GMI-biosensor and further directions are proposed.Entities:
Mesh:
Year: 2005 PMID: 15626616 DOI: 10.1016/j.bios.2004.04.027
Source DB: PubMed Journal: Biosens Bioelectron ISSN: 0956-5663 Impact factor: 10.618