Accuracy and utility of a 10-test disk blood glucose meter.

David M Kendall1, Roy A Kaplan, Carolyn F Paulson, Joan Lee Parkes, Ann M Tideman.   

Abstract

We evaluated the clinical accuracy, precision, and ease-of-use of a whole blood referenced glucose meter system that uses a 10-test disk (Ascensia Confirm Blood Glucose Monitoring System, Bayer Healthcare LLC). The meter system was tested by 100 subjects and eight health care professionals at two separate diabetes centers. Meter blood glucose test results, obtained by the subjects and health care professionals, were accurate and correlated well when compared with laboratory results. The means of the subject and healthcare professional blood glucose results were within 4.8% of the laboratory mean glucose result. When compared with laboratory results, the correlation coefficient was 0.96 for subject meter results and 0.97 for health care professional meter results. Error grid analyses demonstrated that all subject and health care professional blood glucose measurements fell within zones A and B ('no effect on clinical action' and 'altered clinical action with little or no effect on clinical outcome,' respectively). Ninety-three percent (93%) of subjects rated the meter system favorably on an 'ease-of-use' questionnaire. A majority of subjects correctly performed blood glucose self-monitoring tasks simply by reviewing the user guide. In summary, this unique meter that uses a 10-test disk was shown to be both accurate and precise, and subjects with diabetes were able to use the system properly with minimal instructions.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15620431     DOI: 10.1016/j.diabres.2004.05.005

Source DB:  PubMed          Journal:  Diabetes Res Clin Pract        ISSN: 0168-8227            Impact factor:   5.602


  2 in total

Review 1.  Accuracy of point-of-care glucose measurements.

Authors:  Annette Rebel; Mark A Rice; Brenda G Fahy
Journal:  J Diabetes Sci Technol       Date:  2012-03-01

2.  Variable classifications of glycemic index determined by glucose meters.

Authors:  Meng-Hsueh Amanda Lin; Ming-Chang Wu; Jenshinn Lin
Journal:  J Clin Biochem Nutr       Date:  2010-04-29       Impact factor: 3.114

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.