Effectiveness of composite resin polymerization using light-emitting diodes (LEDs) or halogen-based light-curing units.

Bianca Micali1, Roberta Tarkany Basting.   

Abstract

The clinical performance of composite resins is greatly influenced by the quality of the light-curing unit used. The aim of this study was to compare the efficiency of a commercial light-emitting diode (LED) with that of a halogen-based light-curing unit by means of dye penetration of a micro hybrid composite resin. The composite resin evaluated was Filtek Z250 (3M Dental). The composite was filled into acrylic moulds that were randomly polymerized for 40 seconds by each of the light-emitting systems: light-emitting diode Ultraled (Dabi Atlante) or halogen light Degulux (Degussa Hülls) curing units. Immediately after polymerization, each specimen was individually immersed in 1 ml of 2% methylene blue solution at 37 degrees C +/- 2 degrees C. After 24 hours, the specimens were rinsed under running distilled water for 1 minute and stored at 37 degrees C +/- 2 degrees C at relative humidity for 24 hours. The composite resins were removed from the moulds and individually triturated before being immersed in new test tubes containing 1 ml of absolute alcohol for 24 hours. The solutions were filtered and centrifuged for 3 minutes at 4,000 rpm and the supernatant was used to determine absorbance in a spectrophotometer at 590 nm. To verify the differences between groups polymerized by LED or halogen light t-test was applied. No significant differences were found between composite resins light-cured by LED or halogen light-curing unit (p > 0.05). The commercially LED-based light-curing unit is as effective to polymerize hybrid composite resins as the halogen-based unit.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15619883     DOI: 10.1590/s1806-83242004000300016

Source DB:  PubMed          Journal:  Braz Oral Res        ISSN: 1806-8324


  6 in total

1.  Spectrophotometric and colorimetric evaluation of staining of the light cured composite after exposure with different intensities of light curing units.

Authors:  Veeramachaneni Chandrasekhar; L Pramod Reddy; T Jaya Prakash; G Anitha Rao; M Pradeep
Journal:  J Conserv Dent       Date:  2011-10

2.  Stain susceptibility of composite and ceramic CAD/CAM blocks versus direct resin composites with different resinous matrices.

Authors:  Amal Alharbi; Stefano Ardu; Tissiana Bortolotto; Ivo Krejci
Journal:  Odontology       Date:  2016-07-25       Impact factor: 2.634

3.  Effect of Additional Light Curing on Colour Stability of Composite Resins.

Authors:  Kubra Alan Unsal; Emel Karaman
Journal:  Int Dent J       Date:  2021-08-16       Impact factor: 2.607

4.  Composite resin color stability: influence of light sources and immersion media.

Authors:  Patricia Aleixo dos Santos Domingos; Patrícia Petromilli Nordi Sasso Garcia; Ana Luisa Botta Martins de Oliveira; Regina Guenka Palma-Dibb
Journal:  J Appl Oral Sci       Date:  2011 May-Jun       Impact factor: 2.698

5.  Effect of preheat repetition on color stability of methacrylate- and silorane-based composite resins.

Authors:  Mehdi Abed Kahnamouei; Sarah Gholizadeh; Sahand Rikhtegaran; Mehdi Daneshpooy; Soodabeh Kimyai; Parnian Alizadeh Oskoee; Yashar Rezaei
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-12-13

6.  Evaluation of Polymerization Efficacy in Composite Resins via FT-IR Spectroscopy and Vickers Microhardness Test.

Authors:  Tahereh-Sadat Jafarzadeh; Mohammad Erfan; Marjan Behroozibakhsh; Mostafa Fatemi; Reza Masaeli; Yashar Rezaei; Hossein Bagheri; Yasaman Erfan
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2015-12-30
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.