Form analysis using digital signal processing reliably discriminates far-field R waves from P waves.

Norbert M Van Hemel1, Peter Wohlgemuth, Jos G Engbers, Thomas Lawo, Jan Nebaznivy, Milos Taborsky, Joachim Witte, Wim Boute, Dave Munneke, Chris Van Groeningen.   

Abstract

The correct detection of atrial arrhythmias by pacemakers is often limited by the presence of far-field R waves (FFRWs) in the atrial electrogram. Digital signal processing (DSP) of intracardiac signals is assumed to provide improved discrimination between P waves and FFRWs when compared to current methods. For this purpose, 100 bipolar and unipolar intracardiac atrial recordings from 31 patients were collected during pacemaker replacement and used for the off-line application of a novel DSP algorithm. Digital processing of the atrial intracardiac electrogram (IEGM) signals (8 bit, 800 samples/s) included filtering and calculation of the maximum amplitude and slope of the detected events. The form parameter was calculated, being the sum of the most negative value of the amplitude and that of the slope of the detected event. The algorithm collects form parameter data of P waves and FFRWs and composes histograms of these data. A sufficiently large gap between the FFRW and P wave histograms allows discrimination of these two signals based on form parameters. Three independent observers reviewed the reliability of classification with this algorithm. Sensitivity and specificity of FFRW detection were 99.63% and 100%, respectively, and no P waves were falsely classified. It can be concluded that this novel DSP algorithm shows excellent discrimination of FFRWs under off-line conditions and justify the implementation of this algorithm in future pacemakers for real-time discrimination between P waves and FFRWs. This method prevents false mode switching and allows correct and immediate intervention pacing for atrial tachyarrhythmias.

Entities:  

Mesh:

Year:  2004        PMID: 15613124     DOI: 10.1111/j.1540-8159.2004.00694.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  4 in total

1.  Morphological analysis of sinus and retrograde atrial waves detected through a permanent pacemaker atrial lead.

Authors:  Sami Pakarinen; Anne-Mari Vitikainen; Giorgio Corbucci; Lauri Toivonen
Journal:  J Interv Card Electrophysiol       Date:  2010-10-13       Impact factor: 1.900

2.  Clinical application of pacemakers in atrial tachyarrhythmias.

Authors:  W G de Voogt; N M van Hemel
Journal:  Neth Heart J       Date:  2008-10       Impact factor: 2.380

Review 3.  Digital technology in cardiac pacing: methods for morphology analysis of sensed endocavitary signals.

Authors:  Luigi Padeletti; Antonio Michelucci; Gerd Fröhlig; Giorgio Corbucci; Geeske van Oort; Serge S Barold
Journal:  J Interv Card Electrophysiol       Date:  2005-10       Impact factor: 1.900

4.  Diagnostic tools for atrial tachyarrhythmias in implantable pacemakers: a review of technical options and pitfalls.

Authors:  W G de Voogt; N M van Hemel
Journal:  Neth Heart J       Date:  2008-06       Impact factor: 2.380

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.