Clinical results with an ePTFE inflow conduit for mechanical circulatory support.

Tofy Mussivand1, Roland Hetzer, Ettore Vitali, Bart Meyns, Philippe Noirhomme, Reiner Koerfer, Aly El-Banayosy, Ernst Wolner, Georg Wieselthaler, Bruno Reichart, Peter Uberfuhr, Robert Halfmann, Peer Portner.   

Abstract

BACKGROUND: Neurologic complication is an adverse event associated with mechanical circulatory support. To decrease the incidence of embolic cerebrovascular accidents (CVA) during support with the Novacor left ventricular assist system (LVAS), an expanded polytetrafluoroethylene (ePTFE) inflow conduit has been developed and introduced clinically.
METHODS: Using clinical data from Europe and Canada, we conducted a retrospective analysis of the incidence of embolic CVA with the ePTFE inflow conduit (n=88) in comparison with the previously used polyester inflow conduits (n=310, including Vascutek n=155 and Cooley n=155). We calculated freedom from embolic CVA, risk reduction for embolic CVA, and linearized rates of embolic CVA.
RESULTS: A significant decrease in the incidence of embolic CVA was demonstrated with the ePTFE conduit (ePTFE 10% vs polyester 23%, p=0.002). Kaplan-Meier analysis of freedom from embolic CVA at 180 days after implantation was 86% for the ePTFE group vs 72% for the polyester group (log-rank test, 0.0185). We also found an associated risk reduction of 55% in CVA occurrence in the ePTFE group when compared with the Polyester group (hazard ratio, 0.445; 95% confidence limit, 0.222-0.890; p=0.0221). Linearized CVA rates also were decreased at all time intervals after implantation in the ePTFE group.
CONCLUSIONS: Preliminary clinical results with the newly introduced ePTFE inflow conduit provide compelling evidence that the ePTFE conduit material significantly decreases thromboembolic complications during mechanical circulatory support with the Novacor LVAS.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15607665     DOI: 10.1016/j.healun.2003.09.029

Source DB:  PubMed          Journal:  J Heart Lung Transplant        ISSN: 1053-2498            Impact factor:   10.247


  2 in total

1.  Neointima-inducing inflow cannula with titanium mesh for left ventricular assist device.

Authors:  Yukiko Yamada; Tomohiro Nishinaka; Toshihide Mizuno; Yoshiyuki Taenaka; Eisuke Tatsumi; Kenji Yamazaki
Journal:  J Artif Organs       Date:  2011-07-07       Impact factor: 1.731

2.  Platelet activation after implantation of the Levitronix PediVAS in the ovine model.

Authors:  Carl A Johnson; Venkat Shankarraman; Peter D Wearden; Ergin Kocyildirim; Timothy M Maul; John D Marks; J Scott Richardson; Barry N Gellman; Harvey S Borovetz; Kurt A Dasse; William R Wagner
Journal:  ASAIO J       Date:  2011 Nov-Dec       Impact factor: 2.872

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.