Electrically evoked compound action potential amplitude growth functions and HiResolution programming levels in pediatric CII implant subjects.

Marc D Eisen1, Kevin H Franck.   

Abstract

OBJECTIVE: To characterize the amplitude growth functions of the electrically evoked compound action potential (ECAP) in pediatric subjects implanted with the Clarion HiFocus electrode array with respect to electrode position and the presence or absence of a Silastic positioner. Electrophysiologic growth function data are compared with HiResolution (HiRes) psychophysical programming levels.
DESIGN: ECAP growth functions were measured for all electrodes along the implant's array in 16 pediatric subjects. Nine of the patients were implanted with a Silastic positioner, whereas seven had no positioner. ECAP thresholds and growth function slopes were calculated. Fifteen of the 16 patients had psychophysical threshold and maximum comfort levels available. Programming levels and ECAP thresholds were compared within and among the subjects.
RESULTS: ECAP thresholds showed variability among patients, ranging from 178 to 920 nA at 32 musec pulse width. ECAP thresholds did not depend on electrode position along the cochlea but were lower in the presence of the Silastic positioner (p < 0.001). Thresholds determined with the masker-probe versus the alternating polarity paradigms revealed moderate (r = 0.76) correlation. Growth function slopes also showed considerable variation among patients. Unlike thresholds, slopes decreased from apical to basal cochlear locations (p < 0.001) but showed no difference between the absence and presence of the positioner. Programming levels in HiRes were correlated with ECAP threshold levels. When ECAP thresholds were adjusted for each patient by the difference between M level and ECAP threshold at electrode 9, however, overall correlation between the two measurements was excellent (r = 0.98, N = 224).
CONCLUSIONS: In pediatric subjects with the Clarion HiFocus electrode, ECAP growth function thresholds appear to decrease with the presence of the Silastic positioner but are unaffected by electrode position along the array. Growth function slope, however, depends on electrode position along the array but not on the presence of the positioner. ECAP programming levels can reliably predict stimulus intensities within the patients' dynamic ranges, but considerable variability is seen between ECAP thresholds and HiRes programming levels.

Entities:  

Mesh:

Year:  2004        PMID: 15604914     DOI: 10.1097/00003446-200412000-00002

Source DB:  PubMed          Journal:  Ear Hear        ISSN: 0196-0202            Impact factor:   3.570


  23 in total

1.  Intra-Operative Neural Response Telemetry and Acoustic Reflex Assessment using an Advance-In-Stylet Technique and Modiolus-Hugging: A prospective cohort study.

Authors:  Abdulrahman Hagr
Journal:  Sultan Qaboos Univ Med J       Date:  2011-08-15

Review 2.  [Intra- and postoperative electrophysiological diagnostics].

Authors:  T Wesarg; S Arndt; A Aschendorff; R Laszig; R Beck; L Jung; S Zirn
Journal:  HNO       Date:  2017-04       Impact factor: 1.284

3.  Spatial channel interactions in cochlear implants.

Authors:  Qing Tang; Raul Benítez; Fan-Gang Zeng
Journal:  J Neural Eng       Date:  2011-07-13       Impact factor: 5.379

4.  Effects of Stimulus Polarity and Artifact Reduction Method on the Electrically Evoked Compound Action Potential.

Authors:  Michelle L Hughes; Jenny L Goehring; Jacquelyn L Baudhuin
Journal:  Ear Hear       Date:  2017 May/Jun       Impact factor: 3.570

5.  Cochlear Histopathology as Observed in Two Patients With a Cochlear Implant Electrode With Positioner.

Authors:  Takefumi Kamakura; Joseph B Nadol
Journal:  Otol Neurotol       Date:  2016-07       Impact factor: 2.311

6.  Forward Masking in Cochlear Implant Users: Electrophysiological and Psychophysical Data Using Pulse Train Maskers.

Authors:  Youssef Adel; Gaston Hilkhuysen; Arnaud Noreña; Yves Cazals; Stéphane Roman; Olivier Macherey
Journal:  J Assoc Res Otolaryngol       Date:  2017-02-21

7.  A Comparison of Alternating Polarity and Forward Masking Artifact-Reduction Methods to Resolve the Electrically Evoked Compound Action Potential.

Authors:  Jacquelyn L Baudhuin; Michelle L Hughes; Jenny L Goehring
Journal:  Ear Hear       Date:  2016 Jul-Aug       Impact factor: 3.570

8.  Recommendations for Measuring the Electrically Evoked Compound Action Potential in Children With Cochlear Nerve Deficiency.

Authors:  Shuman He; Xiuhua Chao; Ruijie Wang; Jianfen Luo; Lei Xu; Holly F B Teagle; Lisa R Park; Kevin D Brown; Michelle Shannon; Cynthia Warner; Angela Pellittieri; William J Riggs
Journal:  Ear Hear       Date:  2020 May/Jun       Impact factor: 3.570

9.  Comparison of electrically evoked compound action potential thresholds and loudness estimates for the stimuli used to program the Advanced Bionics cochlear implant.

Authors:  Eun Kyung Jeon; Carolyn J Brown; Christine P Etler; Sara O'Brien; Li-Kuei Chiou; Paul J Abbas
Journal:  J Am Acad Audiol       Date:  2010-01       Impact factor: 1.664

10.  Histopathology of the Clarion cochlear implant electrode positioner in a human subject.

Authors:  M Seyyedi; B J Burgess; D K Eddington; B J Gantz; J B Nadol
Journal:  Audiol Neurootol       Date:  2013-06-14       Impact factor: 1.854

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.