Time course of inhaled anaesthetic drug delivery using a new multifunctional closed-circuit anaesthesia ventilator. In vitro comparison with a classical anaesthesia machine.

M M R F Struys1, A F Kalmar, L E C De Baerdemaeker, E P Mortier, G Rolly, J Manigel, W Buschke.   

Abstract

BACKGROUND: The aim of this study was to detail the time-course, defined as the changes in end-tidal drug concentration with time, and consumption of inhaled anaesthetics when using a multifunctional closed-circuit anaesthesia machine in various drug delivery modes, and to compare it with a classical anaesthesia machine using an out-of-circle vaporizer under high and low fresh gas flow conditions.
METHODS: Using an artificial test lung, sevoflurane and desflurane time-course and consumption were compared when using the Zeus apparatus (Dräger, Lubeck, Germany) with direct injection of inhaled anaesthetics or the Primus apparatus (Dräger, Lubeck, Germany) using a classical out-of-circle vaporizer. Anaesthetics were targeted at 1 and 2 MAC end-tidal during 15 min. For both apparatus, out-of-circle high and low fresh gas control (FGC) and for Zeus, auto-control (AC) modes (fixed fresh gas flow at 6 and 1 litre min(-1) and uptake mode) were compared. Time to reach target, initial overshoot and stability at target, and wash-out times were compared.
RESULTS: In FGC, an initial overshoot in end-tidal drug concentration is seen when using 6 litre min(-1) fresh gas flow and a slower time course is observed when using only 1 litre min(-1) in both apparatus. In auto-control mode, the time course of both sevoflurane and desflurane was very fast and not influenced by the changes in fresh gas flow. No overshoot at target was seen. At all settings, the wash-out times were faster when using Zeus than Primus. Inhaled anaesthetic consumption was lowest with the Zeus ventilator in uptake AC mode.
CONCLUSION: A combination of the fastest time course and lowest consumption of sevoflurane and desflurane was found when using the Zeus apparatus in AC uptake mode.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15591326     DOI: 10.1093/bja/aei051

Source DB:  PubMed          Journal:  Br J Anaesth        ISSN: 0007-0912            Impact factor:   9.166


  9 in total

1.  End-tidal versus manually-controlled low-flow anaesthesia.

Authors:  Umberto Lucangelo; Giuliana Garufi; Emanuele Marras; Massimo Ferluga; Federica Turchet; Francesca Bernabè; Lucia Comuzzi; Giorgio Berlot; Walter A Zin
Journal:  J Clin Monit Comput       Date:  2013-10-11       Impact factor: 2.502

2.  Use of a human patient simulator for apnea studies: a preliminary in vitro trial.

Authors:  Debendra Kumar Tripathy; Mridul Dhar; Bharat Bhushan Bhardwaj; K Hemanthkumar; Praveen Talawar; Shalinee Rao
Journal:  Korean J Anesthesiol       Date:  2022-07-13

3.  Mechanical ventilator for delivery of ¹⁷O₂ in brief pulses.

Authors:  J E Baumgardner; E A Mellon; D R Tailor; K Mallikarjunarao; A Borthakur; R Reddy
Journal:  Open Biomed Eng J       Date:  2008-09-24

4.  Optimizing target control of the vessel rich group with volatile anesthetics.

Authors:  Christopher W Connor
Journal:  J Clin Monit Comput       Date:  2018-06-21       Impact factor: 2.502

5.  Cost analysis of two anaesthetic machines: "Primus®" and "Zeus®".

Authors:  Jose Hinz; Nadine Rieske; Bernd Schwien; Aron F Popov; Prashant N Mohite; Oliver Radke; Armin Bartsch; Michael Quintel; Klaus Züchner
Journal:  BMC Res Notes       Date:  2012-01-04

6.  Wash-in and wash-out of sevoflurane in a test-lung model: A comparison between Aisys and FLOW-i.

Authors:  Petter Jakobsson; Madleine Lindgren; Jan G Jakobsson
Journal:  F1000Res       Date:  2017-03-29

7.  Desflurane Consumption During Automated Closed-circuit Delivery is Higher Than When a Conventional Anesthesia Machine is Used With a Simple Vaporizer-O₂-N₂O Fresh Gas Flow Sequence.

Authors:  Sofie De Cooman; Nathalie De Mey; Bram Bc Dewulf; Rik Carette; Thierry Deloof; Maurice Sosnowski; Andre M De Wolf; Jan Fa Hendrickx
Journal:  BMC Anesthesiol       Date:  2008-07-17       Impact factor: 2.217

8.  Cost efficiency of target-controlled inhalational anesthesia.

Authors:  Meenoti Pramod Potdar; Laxmi L Kamat; Manjeet P Save
Journal:  J Anaesthesiol Clin Pharmacol       Date:  2014-04

9.  Agent consumption with the Zeus® in the automated closed circuit anesthesia mode with O2/air mixtures.

Authors:  Sofie De Cooman; Jan F A Hendrickx; Philip John Peyton; Jean-Luc Demeere; Andre M De Wolf
Journal:  BMC Res Notes       Date:  2014-07-23
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.