Effects of ground loop currents on signal intensities in electrospray mass spectrometry.

Richard A Ochran1, Lars Konermann.   

Abstract

The occurrence of electrochemical processes during the operation of an electrospray ionization (ESI) source is well established. In the positive ion mode, electrons are drawn from the ESI metal capillary to a high voltage power supply. These electrons are the product of charge-balancing oxidation reactions taking place at the liquid/metal interface of the ion source. In a recent study, (Anal. Chem.2001, 73, 4836-4844), our group has shown that the introduction of a ground loop can dramatically enhance the rate of these oxidation processes. Such a ground loop can be introduced by connecting the sample infusion syringe (or the liquid chromatography column, in the case of LC-MS studies) to ground. The magnitude of the ground loop current can be controlled by the electrolyte concentration in the analyte solution, and by the dimensions of the capillary connecting the syringe needle and the ESI source. Using ferrocene as a model system, it is demonstrated that the introduction of such a ground loop can significantly enhance the signal intensity of analytes that form electrochemically ionized species during ESI. However, analytes that form protonated molecular ions, such as reserpine, also show higher signal intensities when a ground loop is introduced into the system. This latter observation is attributed to the occurrence of electrolytic solvent (acetonitrile and/or water) oxidation processes. These reactions generate protons within the ion source, and thus facilitate the formation of [M + nH](n+) ions. Overall, this work provides an example of how the careful control of electrochemical parameters can be exploited to optimize signal intensities in ESI-MS.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15589752     DOI: 10.1016/j.jasms.2004.08.005

Source DB:  PubMed          Journal:  J Am Soc Mass Spectrom        ISSN: 1044-0305            Impact factor:   3.109


  15 in total

1.  Electrical equivalence of electrospray ionization with conducting and nonconducting needles.

Authors:  G S Jackson; C G Enke
Journal:  Anal Chem       Date:  1999-09-01       Impact factor: 6.986

2.  Detection and selective dissociation of intact ribosomes in a mass spectrometer.

Authors:  A A Rostom; P Fucini; D R Benjamin; R Juenemann; K H Nierhaus; F U Hartl; C M Dobson; C V Robinson
Journal:  Proc Natl Acad Sci U S A       Date:  2000-05-09       Impact factor: 11.205

3.  Electrochemical processes in electrospray ionization mass spectrometry

Authors: 
Journal:  J Mass Spectrom       Date:  2000-08       Impact factor: 1.982

Review 4.  Protein-folding kinetics and mechanisms studied by pulse-labeling and mass spectrometry.

Authors:  Lars Konermann; Douglas A Simmons
Journal:  Mass Spectrom Rev       Date:  2003 Jan-Feb       Impact factor: 10.946

5.  Mapping of potential gradients within the electrospray emitter.

Authors:  Yan Li; Boguslaw P Pozniak; Richard B Cole
Journal:  Anal Chem       Date:  2003-12-15       Impact factor: 6.986

Review 6.  Studies of biomolecular conformations and conformational dynamics by mass spectrometry.

Authors:  Igor A Kaltashov; Stephen J Eyles
Journal:  Mass Spectrom Rev       Date:  2002 Jan-Feb       Impact factor: 10.946

7.  Quantification and rapid metabolite identification in drug discovery using API time-of-flight LC/MS.

Authors:  N Zhang; S T Fountain; H Bi; D T Rossi
Journal:  Anal Chem       Date:  2000-02-15       Impact factor: 6.986

8.  Electrospray wings for molecular elephants (Nobel lecture).

Authors:  John B Fenn
Journal:  Angew Chem Int Ed Engl       Date:  2003-08-25       Impact factor: 15.336

9.  Electrochemically induced pH changes resulting in protein unfolding in the ion source of an electrospray mass spectrometer.

Authors:  L Konermann; E A Silva; O F Sogbein
Journal:  Anal Chem       Date:  2001-10-15       Impact factor: 6.986

10.  Electrochemical oxidation and cleavage of peptides analyzed with on-line mass spectrometric detection.

Authors:  Hjalmar P Permentier; Ulrik Jurva; Begoña Barroso; Andries P Bruins
Journal:  Rapid Commun Mass Spectrom       Date:  2003       Impact factor: 2.419

View more
  2 in total

1.  2-Nitrophenyl Aryl Sulfides Undergo Both Intramolecular and Electrospray-Induced Intermolecular Oxidation of Sulfur: An Experimental and Theoretical Case Study.

Authors:  Joseph T Moolayil; M George; Daryl Giblin; Michael L Gross
Journal:  Int J Mass Spectrom       Date:  2009-06-01       Impact factor: 1.986

Review 2.  Mass spectrometric methods for monitoring redox processes in electrochemical cells.

Authors:  Herbert Oberacher; Florian Pitterl; Robert Erb; Sabine Plattner
Journal:  Mass Spectrom Rev       Date:  2013-12-10       Impact factor: 10.946

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.