PURPOSE: This study evaluated the feasibility of fiberoptic CO2 laser soldering for the repair of esophageal injuries under tight temperature feedback control in an animal model. Healing was compared to conventional suture closure.
MATERIALS AND METHODS: A CO2 soldering system equipped with infrared transmitting silver halide fibers was used. The soldered tissue temperature was monitored continuously, and laser power was adjusted to provide constant temperature. The procedure was done with 50% bovine serum albumin solder. Longitudinal incisions measuring 8 to 10 mm were made under general anesthesia in the cervical esophagus of 25 rats. Twenty rats (group I) underwent laser tissue bonding; 8 of which were tested in a preliminary study to determine optimal laser parameters. In the remaining 5 rats (group II, controls), closure was performed with 1 layer of 6/0 Vicryl sutures. The rats were sacrificed 2, 3, 4, and 6 weeks postoperatively, and the esophagus was examined histologically.
RESULTS: Optimal temperature was found to be 65 to 70 degrees C and optimal exposure time, 150 to 200 seconds. Laser soldering was successful in 9 of the 12 rats (75%) treated under optimal settings; suturing was successful in 4 of the 5 control rats (80%). There were no significant differences between the groups in healing or complication rates.
CONCLUSIONS: These results indicate that the CO2 laser soldering technique is a valid option for the correction of esophageal tears or incisions in rats. The confirmation and extension of these findings in further studies with larger animals may ultimately lead to the routine in vivo use of temperature-controlled laser repair for the esophagus and other organs. This method lends itself to endoscopic bonding of tissues.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15577780     DOI: 10.1016/j.otohns.2004.08.002

Source DB:  PubMed          Journal:  Otolaryngol Head Neck Surg        ISSN: 0194-5998            Impact factor:   3.497


  3 in total

1.  The effects of ankaferd blood stopper on the recovery process in an experimental oesophageal perforation model.

Authors:  Necla Gürbüz Sarıkaş; Tanzer Korkmaz; Nurettin Kahramansoy; Ali Kılıçgün; Çetin Boran; Güledal Boztaş
Journal:  Balkan Med J       Date:  2015-01-01       Impact factor: 2.021

2.  Temperature-controlled laser-soldering system and its clinical application for bonding skin incisions.

Authors:  David Simhon; Ilan Gabay; Gregory Shpolyansky; Tamar Vasilyev; Israel Nur; Roberto Meidler; Ossama Abu Hatoum; Abraham Katzir; Moshe Hashmonai; Doron Kopelman
Journal:  J Biomed Opt       Date:  2015       Impact factor: 3.170

3.  Evaluation of Laser Tissue Welding and Laser-Tissue Soldering for Mucosal and Vascular Repair.

Authors:  Yusuf Abbas Mistry; Srivalli S Natarajan; Suraj A Ahuja
Journal:  Ann Maxillofac Surg       Date:  2018 Jan-Jun
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.