Literature DB >> 15570408

Corrosion of orthodontic pliers using different sterilization procedures.

Andrea Wichelhaus1, Gert Brauchle, Matthias Mertmann, Franz Günter Sander.   

Abstract

AIM: The aim of this study was to investigate the corrosion resistance of orthodontic pliers from several manufacturers following different sterilization procedures, namely heat sterilization and the use of a disinfectant agent (Sekusept Extra N in an ultrasonic bath.
MATERIAL AND METHODS: Employed in this study were ten distal-end cutters and ten Weingart pliers manufactured by Aesculap, ETM and Hu-Friedy, examined under electron microscopy for evidence of corrosion following 500 sterilization cycles. The corrosion data were recorded photographically and graphically. Metallographic microsections were prepared from each pair of pliers and were then analyzed by means of EDX measurements to assess the chemical alloy composition.
RESULTS: The light and electron microscopic evaluation showed that both heat sterilization and high level disinfection caused corrosive changes on the pliers. However, the type of corrosion differed between the two sterilization methods. The main type of corrosion with Sekusept Extra N was pitting, while with heat sterilization it was surface corrosion. Heat sterilization was the type of corrosion that caused the most corrosive changes, regardless of which type of corrosion was considered. However, these results must be viewed critically, because pitting corrosion is obviously a more severe problem than surface corrosion. In comparing the pliers of various manufacturers, differences were noted. However, it was difficult to rank them when the correct maintenance regime was adhered to. The soldered joint gap was revealed to be a specific weak point, and that area proved to be inadequate due to qualities related to their production, particularly in the Hu-Friedy pliers.
CONCLUSIONS: This study showed that heat sterilization leads to less corrosion than cold disinfection. Corroded pliers can be restored to a useable condition by re-polishing, though it is important that the instructions for their care be strictly adhered to.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15570408     DOI: 10.1007/s00056-004-0417-9

Source DB:  PubMed          Journal:  J Orofac Orthop        ISSN: 1434-5293            Impact factor:   1.938


  3 in total

1.  In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study.

Authors:  Livia Kelly Ferraz Nunes; Érika Lopes Carneiro; Nathalia Barbosa Palomares; Lilian Senna; Helio Sampaio-Filho; José Augusto Mendes Miguel
Journal:  Dental Press J Orthod       Date:  2021-10-25

2.  Corrosion in Haas expanders with and without use of an antimicrobial agent: an in situ study.

Authors:  Cristhiane Ristum Bagatin; Izabel Yoko Ito; Marcela Cristina Damião Andrucioli; Paulo Nelson-Filho; José Tarcísio Lima Ferreira
Journal:  J Appl Oral Sci       Date:  2011 Nov-Dec       Impact factor: 2.698

3.  Biosafety conducts adopted by orthodontists.

Authors:  Camila Gonçalves Jezini Monteiro; Mariana Martins E Martins; Adriana de Alcantara Cury-Saramago; Henry Pinheiro Teixeira
Journal:  Dental Press J Orthod       Date:  2018 May-Jun
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.