Guidelines on preparation, certification, and use of certified plasmas for ISI calibration and INR determination.

A M H P van den Besselaar1, T W Barrowcliffe, L L Houbouyan-Réveillard, J Jespersen, M Johnston, L Poller, A Tripodi.   

Abstract

Reliable international normalized ratio (INR) determination depends on accurate values for international sensitivity index (ISI) and mean normal prothrombin time (MNPT). Local ISI calibration can be performed to obtain reliable INR. Alternatively, the laboratory may determine INR directly from a line relating local log(prothrombin time [PT]) to log(INR). This can be done by means of lyophilized or frozen plasmas to which certified values of PT or INR have been assigned. Currently there is one procedure for local calibration with certified plasmas which is a modification of the WHO method of ISI determination. In the other procedure, named 'direct' INR determination, certified plasmas are used to calculate a line relating log(PT) to log(INR). The number of certified plasmas for each procedure depends on the method of preparation and type of plasma. Lyophilization of plasma may induce variable effects on the INR, the magnitude of which depends on the type of thromboplastin used. Consequently, the manufacturer or supplier of certified plasmas must assign the values for different (reference) thromboplastins and validate the procedure for reliable ISI calibration or 'direct' INR determination. Certification of plasmas should be performed by at least three laboratories. Multiple values should be assigned if the differences between thromboplastin systems are greater than 10%. Testing of certified plasmas for ISI calibration may be performed in quadruplicate in the same working session. It is recommended to repeat the measurements on three sessions or days to control day-to-day variation. Testing of certified plasmas for 'direct' INR determination should be performed in at least three sessions or days. Correlation lines for ISI calibration and for 'direct' INR determination should be calculated by means of orthogonal regression. Quality assessment of the INR with certified plasmas should be performed regularly and should be repeated whenever there is a change in reagent batch or in instrument. Discrepant results obtained by users of certified plasmas should be reported to manufacturers or suppliers.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15550026     DOI: 10.1111/j.1538-7836.2004.00970.x

Source DB:  PubMed          Journal:  J Thromb Haemost        ISSN: 1538-7836            Impact factor:   5.824


  11 in total

1.  Kinetic laws, phase-phase expansions, renormalization group, and INR calibration.

Authors:  Marcel O Vlad; Alexandru D Corlan; Federico Morán; Rainer Spang; Peter Oefner; John Ross
Journal:  Proc Natl Acad Sci U S A       Date:  2009-04-06       Impact factor: 11.205

2.  The need for an adapted initiation nomogram during Fiix prothrombin time monitoring of warfarin.

Authors:  Pall T Onundarson; Brynja R Gudmundsdottir
Journal:  J Thromb Thrombolysis       Date:  2019-11       Impact factor: 2.300

3.  Clinical usefulness of international normalized ratio calibration of prothrombin time in patients with chronic liver disease.

Authors:  Jun Hyung Lee; Oh Joo Kweon; Mi-Kyung Lee; Hyun Woong Lee; Hyung Joon Kim; Hye Ryoun Kim
Journal:  Int J Hematol       Date:  2015-06-12       Impact factor: 2.490

4.  Effect of anticoagulant adjustment on prothrombin time test using two different PT reagents in patients with elevated hematocrit.

Authors:  Vanessa M Silva; Denise C Rezende; Edna S Garcia; Cyrillo Cavalheiro; Celia C Strunz
Journal:  Pract Lab Med       Date:  2020-10-17

5.  The international normalized ratio of prothrombin time in the model for end-stage liver disease score: a reliable measure.

Authors:  Patrick S Kamath; W Ray Kim
Journal:  Clin Liver Dis       Date:  2009-02       Impact factor: 6.126

6.  Results expression for tests used to measure the anticoagulant effect of new oral anticoagulants.

Authors:  Armando Tripodi
Journal:  Thromb J       Date:  2013-06-28

7.  Evaluation of the Steelex M600H coagulometer prothrombin time-international normalized ratio assay with Steelex test reagents.

Authors:  Yasemin U Budak; Kağan Huysal; Murat Polat; Gülsevil Tarakçi; Hakan Uçar
Journal:  Biochem Med (Zagreb)       Date:  2012       Impact factor: 2.313

8.  Paving the way for establishing a reference measurement system for standardization of plasma prothrombin time: Harmonizing the manual tilt tube method.

Authors:  Antonius M H P van den Besselaar; Claudia J J van Rijn; Charmane F Abdoel; Veena Chantarangkul; Erica Scalambrino; Steve Kitchen; Armando Tripodi; Anita M Woolley; Lidia Padovan; Christa M Cobbaert
Journal:  J Thromb Haemost       Date:  2020-06-25       Impact factor: 5.824

Review 9.  Updates on Anticoagulation and Laboratory Tools for Therapy Monitoring of Heparin, Vitamin K Antagonists and Direct Oral Anticoagulants.

Authors:  Osamu Kumano; Kohei Akatsuchi; Jean Amiral
Journal:  Biomedicines       Date:  2021-03-07

10.  Monitoring of anticoagulant therapy in heart disease: considerations for the current assays.

Authors:  Mohammadali Boroumand; Hamidreza Goodarzynejad
Journal:  J Tehran Heart Cent       Date:  2010-05-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.