| Literature DB >> 15542169 |
Ritsu Yamamoto1, Akio Yonesaka, Seiko Nishioka, Hidemichi Watari, Takayuki Hashimoto, Daichi Uchida, Hiroshi Taguchi, Takeshi Nishioka, Kazuo Miyasaka, Noriaki Sakuragi, Hiroki Shirato.
Abstract
The feasibility and accuracy of high dose three-dimensional conformal boost (3DCB) using three internal fiducial markers and a two-orthogonal X-ray set-up of the real-time tumor-tracking system on patients with gynecological malignacy were investigated in 10 patients. The standard deviation of the distribution of systematic deviations (Sigma) was reduced from 3.8, 4.6, and 4.9 mm in the manual set-up to 2.3, 2.3 and 2.7 mm in the set-up using the internal markers. The average standard deviation of the distribution of random deviations (sigma) was reduced from 3.7, 5.0, and 4.5 mm in the manual set-up to 3.3, 3.0, and 4.2 mm in the marker set-up. The appropriate PTV margin was estimated to be 10.2, 12.8, and 12.9 mm in the manual set-up and 6.9, 6.7, and 8.3 mm in the gold marker set-up, respectively, using the formula 2Sigma + 0.7sigma. Set-up of the patients with three markers and two fluoroscopy is useful to reduce PTV margin and perform 3DCB.Entities:
Mesh:
Year: 2004 PMID: 15542169 DOI: 10.1016/j.radonc.2004.08.006
Source DB: PubMed Journal: Radiother Oncol ISSN: 0167-8140 Impact factor: 6.280