Literature DB >> 15539726

Supplemental oxygen impairs detection of hypoventilation by pulse oximetry.

Eugene S Fu1, John B Downs, John W Schweiger, Rafael V Miguel, Robert A Smith.   

Abstract

STUDY
OBJECTIVE: This two-part study was designed to determine the effect of supplemental oxygen on the detection of hypoventilation, evidenced by a decline in oxygen saturation (Spo(2)) with pulse oximetry.
DESIGN: Phase 1 was a prospective, patient-controlled, clinical trial. Phase 2 was a prospective, randomized, clinical trial.
SETTING: Phase 1 took place in the operating room. Phase 2 took place in the postanesthesia care unit (PACU). PATIENTS: In phase 1, 45 patients underwent abdominal, gynecologic, urologic, and lower-extremity vascular operations. In phase 2, 288 patients were recovering from anesthesia.
INTERVENTIONS: In phase 1, modeling of deliberate hypoventilation entailed decreasing by 50% the minute ventilation of patients receiving general anesthesia. Patients breathing a fraction of inspired oxygen (Fio(2)) of 0.21 (n = 25) underwent hypoventilation for up to 5 min. Patients with an Fio(2) of 0.25 (n = 10) or 0.30 (n = 10) underwent hypoventilation for 10 min. In phase 2, spontaneously breathing patients were randomized to breathe room air (n = 155) or to receive supplemental oxygen (n = 133) on arrival in the PACU. MEASUREMENTS AND
RESULTS: In phase 1, end-tidal carbon dioxide and Spo(2) were measured during deliberate hypoventilation. A decrease in Spo(2) occurred only in patients who breathed room air. No decline occurred in patients with Fio(2) levels of 0.25 and 0.30. In phase 2, Spo(2) was recorded every min for up to 40 min in the PACU. Arterial desaturation (Spo(2) < 90%) was fourfold higher in patients who breathed room air than in patients who breathed supplemental oxygen (9.0% vs 2.3%, p = 0.02).
CONCLUSION: Hypoventilation can be detected reliably by pulse oximetry only when patients breathe room air. In patients with spontaneous ventilation, supplemental oxygen often masked the ability to detect abnormalities in respiratory function in the PACU. Without the need for capnography and arterial blood gas analysis, pulse oximetry is a useful tool to assess ventilatory abnormalities, but only in the absence of supplemental inspired oxygen.

Entities:  

Mesh:

Year:  2004        PMID: 15539726     DOI: 10.1378/chest.126.5.1552

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  58 in total

1.  Is it time to change the approach to oxygen therapy in the breathless patient?

Authors:  Richard Beasley; Sarah Aldington; Geoffrey Robinson
Journal:  Thorax       Date:  2007-10       Impact factor: 9.139

2.  A procedure for determining subject-specific pulse oxygen saturation response.

Authors:  Kyle M Burk; Joseph A Orr
Journal:  Med Biol Eng Comput       Date:  2020-01-27       Impact factor: 2.602

3.  Respiratory monitoring: physiological and technical considerations.

Authors:  Daniel E Becker; Andrew B Casabianca
Journal:  Anesth Prog       Date:  2009

4.  The law of unintended consequences: when pain management leads to medication errors.

Authors:  Steven Hanks
Journal:  P T       Date:  2008-07

5.  Capnographic Monitoring in Routine EGD and Colonoscopy With Moderate Sedation: A Prospective, Randomized, Controlled Trial.

Authors:  Paresh P Mehta; Gursimran Kochhar; Mazen Albeldawi; Brian Kirsh; Maged Rizk; Brian Putka; Binu John; Yinghong Wang; Nicole Breslaw; Rocio Lopez; John J Vargo
Journal:  Am J Gastroenterol       Date:  2016-02-23       Impact factor: 10.864

Review 6.  How best to approach endoscopic sedation?

Authors:  Michaela Müller; Till Wehrmann
Journal:  Nat Rev Gastroenterol Hepatol       Date:  2011-07-12       Impact factor: 46.802

7.  Assessment of noninvasive acoustic respiration rate monitoring in patients admitted to an Emergency Department for drug or alcoholic poisoning.

Authors:  Youcef Guechi; Amélie Pichot; Denis Frasca; Fatima Rayeh-Pelardy; Jean-Yves Lardeur; Olivier Mimoz
Journal:  J Clin Monit Comput       Date:  2015-01-23       Impact factor: 2.502

Review 8.  Airway Management During Upper GI Endoscopic Procedures: State of the Art Review.

Authors:  Basavana Goudra; Preet Mohinder Singh
Journal:  Dig Dis Sci       Date:  2016-11-12       Impact factor: 3.199

9.  Evaluation of the SEDline to improve the safety and efficiency of conscious sedation.

Authors:  Thomas D Caputo; Michael A E Ramsay; Jeffrey A Rossmann; M Miles Beach; Garth R Griffiths; Benjamin Meyrat; James B Barnes; David G Kerns; Brad Crump; Barnett Bookatz; Paul Ezzo
Journal:  Proc (Bayl Univ Med Cent)       Date:  2011-07

10.  A comparison of the incidence of hypercapnea in non-obese and morbidly obese peri-operative patients using the SenTec transcutaneous pCO(2) monitor.

Authors:  Roy G Soto; Maurice Davis; Michael J Faulkner
Journal:  J Clin Monit Comput       Date:  2013-11-29       Impact factor: 2.502

View more