Geometrical accuracy of the Novalis stereotactic radiosurgery system for trigeminal neuralgia.

Javad Rahimian1, Joseph C Chen, Ajay A Rao, Michael R Girvigian, Michael J Miller, Hugh E Greathouse.   

Abstract

OBJECT: Stringent geometrical accuracy and precision are required in the stereotactic radiosurgical treatment of patients. Accurate targeting is especially important when treating a patient in a single fraction of a very high radiation dose (90 Gy) to a small target such as that used in the treatment of trigeminal neuralgia (3 to 4-mm diameter). The purpose of this study was to determine the inaccuracies in each step of the procedure including imaging, fusion, treatment planning, and finally the treatment. The authors implemented a detailed quality-assurance program.
METHODS: Overall geometrical accuracy of the Novalis stereotactic system was evaluated using a Radionics Geometric Phantom Chamber. The phantom has several magnetic resonance (MR) and computerized tomography (CT) imaging-friendly objects of various shapes and sizes. Axial 1-mm-thick MR and CT images of the phantom were acquired using a T1-weighted three-dimensional spoiled gradient recalled pulse sequence and the CT scanning protocols used clinically in patients. The absolute errors due to MR image distortion, CT scan resolution, and the image fusion inaccuracies were measured knowing the exact physical dimensions of the objects in the phantom. The isocentric accuracy of the Novalis gantry and the patient support system was measured using the Winston-Lutz test. Because inaccuracies are cumulative, to calculate the system's overall spatial accuracy, the root mean square (RMS) of all the errors was calculated. To validate the accuracy of the technique, a 1.5-mm-diameter spherical marker taped on top of a radiochromic film was fixed parallel to the x-z plane of the stereotactic coordinate system inside the phantom. The marker was defined as a target on the CT images, and seven noncoplanar circular arcs were used to treat the target on the film. The calculated system RMS value was then correlated with the position of the target and the highest density on the radiochromic film. The mean spatial errors due to image fusion and MR imaging were 0.41+/-0.3 and 0.22+/-0.1 mm, respectively. Gantry and couch isocentricities were 0.3+/-0.1 and 0.6+/-0.15 mm, respectively. The system overall RMS values were 0.9 and 0.6 mm with and without the couch errors included, respectively (isocenter variations due to couch rotation are microadjusted between couch positions). The positional verification of the marker was within 0.7+/-0.1 mm of the highest optical density on the radiochromic film, correlating well with the system's overall RMS value. The overall mean system deviation was 0.32+/-0.42 mm.
CONCLUSIONS: The highest spatial errors were caused by image fusion and gantry rotation. A comprehensive quality-assurance program was developed for the authors' stereotactic radiosurgery program that includes medical imaging, linear accelerator mechanical isocentricity, and treatment delivery. For a successful treatment of trigeminal neuralgia with a 4-mm cone, the overall RMS value of equal to or less than 1 mm must be guaranteed.

Entities:  

Mesh:

Year:  2004        PMID: 15537189     DOI: 10.3171/jns.2004.101.supplement 3.0351

Source DB:  PubMed          Journal:  J Neurosurg        ISSN: 0022-3085            Impact factor:   5.115


  17 in total

Review 1.  Target delineation and optimal radiosurgical dose for pituitary tumors.

Authors:  Giuseppe Minniti; Mattia Falchetto Osti; Maximillian Niyazi
Journal:  Radiat Oncol       Date:  2016-10-11       Impact factor: 3.481

2.  A survey of image-guided radiation therapy use in the United States.

Authors:  Daniel R Simpson; Joshua D Lawson; Sameer K Nath; Brent S Rose; Arno J Mundt; Loren K Mell
Journal:  Cancer       Date:  2010-08-15       Impact factor: 6.860

3.  Developing minimally invasive surgery centers within kaiser permanente: the integrated multidisciplinary experience of los angeles.

Authors:  Gary W Chien; Maher A Abbas
Journal:  Perm J       Date:  2009

4.  Helical tomotherapy for single and multiple liver tumours.

Authors:  Tsair-Fwu Lee; Pei-Ju Chao; Fu-Min Fang; Te-Jen Su; Stephen W Leung; Hsuan-Chih Hsu
Journal:  Radiat Oncol       Date:  2010-06-24       Impact factor: 3.481

5.  Targeting accuracy at couch kick for a frameless image guided radiosurgery system.

Authors:  Yimei Huang; Bo Zhao; Joshua Kim; Ning Wen; Indrin J Chetty; Salim Siddiqui
Journal:  J Radiosurg SBRT       Date:  2018

6.  Validation of Pinnacle treatment planning system for use with Novalis delivery unit.

Authors:  Vladimir Faygelman; Dylan Hunt; Luke Walker; Richard Mueller; Mary Lou Demarco; Thomas Dilling; Craig Stevens; Geoffrey Zhang
Journal:  J Appl Clin Med Phys       Date:  2010-06-15       Impact factor: 2.102

7.  Image fusion pitfalls for cranial radiosurgery.

Authors:  Benjamin P Jonker
Journal:  Surg Neurol Int       Date:  2013-04-17

8.  Radiosurgical third ventriculostomy: Technical note.

Authors:  Guillermo Axayacalt Gutiérrez-Aceves; Sergio Moreno-Jiménez; Miguel Ángel Celis; Mariana Hernández-Bojórquez
Journal:  Surg Neurol Int       Date:  2012-10-13

9.  Improvement of therapeutic index for brain tumors with daily image guidance.

Authors:  Lisa Be Shields; James M Coons; Catherine Dedich; Maria Ragains; Kristi Scalf; Todd W Vitaz; Aaron C Spalding
Journal:  Radiat Oncol       Date:  2013-12-02       Impact factor: 3.481

10.  Simulation of intrafraction motion and overall geometrical accuracy of a frameless intracranial radiosurgery process.

Authors:  Vladimir Feygelman; Luke Walker; Prakash Chinnaiyan; Kenneth Forster
Journal:  J Appl Clin Med Phys       Date:  2008-10-24       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.