The feasibility of ST-segment monitoring with a subcutaneous device.

Zhendong Song1, Janice Jenkins, Martin Burke, Robert Arzbaecher.   

Abstract

We have developed a subcutaneous cardiac monitor/alarm for patients at high risk that alerts companions or bystanders if cardiac arrest occurs. The leadless device has 4 electrodes mounted on its corners. An added feature of the implantable monitor/alarm is the ability to detect ST-segment deviations present in acute ischemia. We obtained acute ischemia electrocardiogram (ECG) data of 30 patients from the Utah 192-site body surface maps and determined the left precordial region where our device would be implanted. (The device is implanted at the location where its electrodes yield maximum QRS amplitude.) Unipolar body surface potentials from the chosen area were further interpolated so that closely spaced bipolar electrocardiograms can be derived by performing algebraic subtraction of these "unipolar" ECGs. Results show that the summation of the absolute values of ST-deviations from a pair of diagonal bipolar leads from the optimal subarea (3 x 6 cm rectangle) provides a measure comparable to the maximum ST-deviation among the 4 corner unipolar leads of the same area. This suggests that ST-segment deviations seen at neighboring precordial sites are sufficiently different that significant cancellation of these deviations does not occur in forming close bipolar ECGs. Our conclusion is that during acute ischemia, bipolar ECG recordings between neighboring precordial sites can present ST-segment deviations that are large enough for automatic detection in many patients. Furthermore, the same electrodes used by our subcutaneous device in monitoring R-waves in arrhythmia detection can also be used in monitoring ST-segment deviations for acute ischemia detection.

Entities:  

Mesh:

Year:  2004        PMID: 15534837     DOI: 10.1016/j.jelectrocard.2004.08.053

Source DB:  PubMed          Journal:  J Electrocardiol        ISSN: 0022-0736            Impact factor:   1.438


  3 in total

1.  Finite difference and lead field methods in designing implantable ECG monitor.

Authors:  Juho Väisänen; Jari Hyttinen; Jaakko Malmivuo
Journal:  Med Biol Eng Comput       Date:  2006-08-11       Impact factor: 2.602

Review 2.  Wireless and inductively powered implant for measuring electrocardiogram.

Authors:  Jarno Riistama; Juho Väisänen; Sami Heinisuo; Hanna Harjunpää; Satu Arra; Kati Kokko; Maunu Mäntylä; Jutta Kaihilahti; Pekka Heino; Minna Kellomäki; Outi Vainio; Jukka Vanhala; Jukka Lekkala; Jari Hyttinen
Journal:  Med Biol Eng Comput       Date:  2007-10-11       Impact factor: 2.602

3.  A statistically based acute ischemia detection algorithm suitable for an implantable device.

Authors:  Bruce Hopenfeld; M Sasha John; Tim A Fischell; Steven R Johnson
Journal:  Ann Biomed Eng       Date:  2012-06-28       Impact factor: 3.934