Literature DB >> 15533471

Novel algorithm for improving accuracy of ultrasound measurement of residual urine volume according to bladder shape.

Jae Youn Hwang1, Seok-Soo Byun, Seung-June Oh, Hee Chan Kim.   

Abstract

OBJECTIVES: To suggest a novel algorithm based on individually determined proportionality constants to improve the accuracy of noninvasively calculated residual urine volume using ultrasound bladder images.
METHODS: The residual urine volume of 116 subjects was calculated from ultrasound images obtained using a conventional two-dimensional ultrasound device at two institutions. The proposed method uses individually determined proportionality constants for specific bladder shapes. The corresponding true residual volumes were determined by measuring the catheterized volume. The accuracy of our method was confirmed by comparing the mean and standard deviation of the fractional absolute error (FAE) versus true volume with values obtained using other methods, which rely on fixed constants.
RESULTS: The calculated residual bladder volume of 116 patients correlated significantly (r = 0.94, P <0.001) and the differences from the true values were not statistically significant (P = 0.42). The mean FAE was 0.17 +/- 0.10. Subgroups divided by sex, postvoid residual urine volume (over and under 150 mL), and institution produced different results (mean FAE for men 0.15 +/- 0.11 and for women 0.18 +/- 0.12; mean FAE for greater than 150 mL, 0.16 +/- 0.11 and for less than 150 mL, 0.17 +/- 0.14; mean FAE for Seoul National University Hospital 0.15 +/- 0.10 and for Seoul National University Bundang Hospital 0.18 +/- 0.18). Our method provided better accuracy than other known formulas using a conventional two-dimensional ultrasound scanner. The latter multiply the product of the sagittal height and depth and transverse width by a fixed coefficient.
CONCLUSIONS: Our method, based on individually determined proportionality constants, improved the accuracy of residual bladder volume determinations. The proposed algorithm can easily be implemented on a personal computer connected to a conventional ultrasound imaging unit.

Entities:  

Mesh:

Year:  2004        PMID: 15533471     DOI: 10.1016/j.urology.2004.06.054

Source DB:  PubMed          Journal:  Urology        ISSN: 0090-4295            Impact factor:   2.649


  6 in total

1.  The use of inversion mode and 3D manual segmentation in volume measurement of fetal fluid-filled structures: comparison with Virtual Organ Computer-aided AnaLysis (VOCAL).

Authors:  J P Kusanovic; J K Nien; L F Gonçalves; J Espinoza; W Lee; M Balasubramaniam; E Soto; O Erez; R Romero
Journal:  Ultrasound Obstet Gynecol       Date:  2008-02       Impact factor: 7.299

2.  Determination of postvoid residual by translabial ultrasound.

Authors:  H P Dietz; D Velez; K L Shek; A Martin
Journal:  Int Urogynecol J       Date:  2012-05-10       Impact factor: 2.894

3.  Bladder volume correction factors measured with 3D ultrasound and BladderScan.

Authors:  Naomi N Vinod; Anna S Nagle; Hameeda A Naimi; Hiren Kolli; Derek Sheen; Naveen Nandanan; Laura R Carucci; John E Speich; Adam P Klausner
Journal:  Can J Urol       Date:  2019-08       Impact factor: 1.344

Review 4.  A comprehensive survey on non-invasive wearable bladder volume monitoring systems.

Authors:  Morteza Zakeri Nasrabadi; Hamideh Tabibi; Mahsa Salmani; Mahdieh Torkashvand; Eisa Zarepour
Journal:  Med Biol Eng Comput       Date:  2021-07-14       Impact factor: 2.602

5.  Comparison of 2D and 3D ultrasound methods to measure serial bladder volumes during filling: Steps toward development of non-invasive ultrasound urodynamics.

Authors:  Anna S Nagle; Rachel J Bernardo; Jary Varghese; Laura R Carucci; Adam P Klausner; John E Speich
Journal:  Bladder (San Franc)       Date:  2018-01-04

6.  Evaluation of electrical impedance tomography for determination of urinary bladder volume: comparison with standard ultrasound methods in healthy volunteers.

Authors:  Dorothea Leonhäuser; Carlos Castelar; Thomas Schlebusch; Martin Rohm; Rüdiger Rupp; Steffen Leonhardt; Marian Walter; Joachim O Grosse
Journal:  Biomed Eng Online       Date:  2018-07-13       Impact factor: 2.819