An algorithm to predict implantable cardioverter-defibrillator lead failure.

Bruce D Gunderson1, Amisha S Patel, Chad A Bounds, Richard K Shepard, Mark A Wood, Kenneth A Ellenbogen.   

Abstract

OBJECTIVES: The goal of this analysis was to test an algorithm that identifies implantable cardioverter-defibrillator (ICD) lead problems before clinical failure and/or inappropriate therapy.
BACKGROUND: The ICD lead failures typically present as inappropriate shock therapy. Identifying lead failures before their clinical presentation may prevent patient discomfort, improve device longevity, and avoid device-induced proarrhythmia.
METHODS: We tested an algorithm that uses two measures of oversensing and one measure of abnormal impedance to detect a lead failure. The oversensing measures consisted of a counter for RR intervals <140 ms and nonsustained ventricular tachycardia episodes with mean RR interval <200 ms. The impedance measure tracked lead impedances every day and each week. Abnormal impedance was defined as a decrease in impedances or an outlier value compared with baseline. Lead failures were identified when both oversensing measures were met or abnormal impedance and one oversensing measure occurred. The stored data from 696 patients with an ICD were analyzed to determine the sensitivity and specificity of the algorithm to detect lead failures.
RESULTS: Twenty-nine patients demonstrated clinical lead failures with an average of 6 +/- 9 inappropriate shocks per patient. The two oversensing measures used in the algorithm predicted 72% (21 of 29) of the lead failures. Fulfilling at least two of the three impedance and oversensing measures, the sensitivity of our algorithm was 83% (24 of 29) with a 100% (667 of 667) specificity.
CONCLUSION: Oversensing combined with abnormal impedance trends may be used to identify ICD lead failures with high sensitivity and very high specificity.

Entities:  

Mesh:

Year:  2004        PMID: 15519026     DOI: 10.1016/j.jacc.2004.07.042

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  9 in total

1.  Can we predict and prevent adverse events related to high-voltage implantable cardioverter defibrillator lead failure?

Authors:  Renato Pietro Ricci; Carlo Pignalberi; Barbara Magris; Stefano Aquilani; Vito Altamura; Loredana Morichelli; Antonio Porfili; Laura Quarta; Fabio Saputo; Massimo Santini
Journal:  J Interv Card Electrophysiol       Date:  2011-09-01       Impact factor: 1.900

2.  Focus on management of pacemaker and ICD advisories, recalls, and alerts.

Authors:  Mitesh S Amin; Kenneth A Ellenbogen
Journal:  Curr Treat Options Cardiovasc Med       Date:  2006-09

3.  A prospective comparison of remote monitoring systems in implantable cardiac defibrillators: potential effects of frequency of transmissions.

Authors:  Ermenegildo de Ruvo; Luigi Sciarra; Anna Maria Martino; Marco Rebecchi; Renzo Venanzio Iulianella; Francesco Sebastiani; Alessandro Fagagnini; Alessio Borrelli; Antonio Scarà; Domenico Grieco; Claudia Tota; Federica Stirpe; Leonardo Calò
Journal:  J Interv Card Electrophysiol       Date:  2016-01       Impact factor: 1.900

4.  Performance of an ICD algorithm to detect lead noise and reduce inappropriate shocks.

Authors:  Scott Beau; Stephen Greer; Christopher R Ellis; Jeffrey Keeney; Shubha Asopa; Edith Arnold; Avi Fischer
Journal:  J Interv Card Electrophysiol       Date:  2016-01-06       Impact factor: 1.900

5.  Turning a blind eye to the far field: Are we burying the evidence? A case of abrupt catastrophic implantable cardioverter defibrillator lead failure causing sudden death.

Authors:  Stephen Tuohy; Paul Ryan; Joseph Galvin
Journal:  HeartRhythm Case Rep       Date:  2015-09-04

6.  Inappropriate detection of an intraventricular conduction delay as a lead failure due to an increase in the sensing integrity counter in a patient with a cardiac resynchronization therapy-defibrillator.

Authors:  Kentaro Ozu; Hitoshi Minamiguchi; Shozo Konishi; Hiroya Mizuno; Yasushi Sakata
Journal:  HeartRhythm Case Rep       Date:  2020-10-24

7.  Novel ventricular tachyarrhythmia detection enhancement detects undertreated life-threatening arrhythmias.

Authors:  Bruce L Wilkoff; Laurence D Sterns; Michael S Katcher; Gaurav Upadhyay; Peter Seizer; Chaoyi Kang; Jennifer Rhude; Kevin J Davis; Avi Fischer
Journal:  Heart Rhythm O2       Date:  2021-11-18

8.  Design of the evolution of management strategies of heart failure patients with implantable defibrillators (EVOLVO) study to assess the ability of remote monitoring to treat and triage patients more effectively.

Authors:  Maurizio Marzegalli; Maurizio Landolina; Maurizio Lunati; Giovanni B Perego; Alessia Pappone; Giuseppe Guenzati; Carlo Campana; Maria Frigerio; Gianfranco Parati; Antonio Curnis; Irene Colangelo; Sergio Valsecchi
Journal:  Trials       Date:  2009-06-18       Impact factor: 2.279

9.  Patient alerting features in implantable defibrillators.

Authors:  Dirk Vollmann; Markus Zabel; Johnson Francis
Journal:  Indian Pacing Electrophysiol J       Date:  2008-02-01
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.