Performance of an ICD algorithm to detect lead noise and reduce inappropriate shocks.

Scott Beau1, Stephen Greer2, Christopher R Ellis3, Jeffrey Keeney4, Shubha Asopa4, Edith Arnold4, Avi Fischer4.   

Abstract

BACKGROUND: Implantable cardioverter-defibrillators (ICD) provide treatment for life-threatening ventricular tachyarrhythmias. Failure of the pace/sense conductor of an ICD lead can cause noise on the sensing electrogram (EGM) that may be misinterpreted as ventricular activity, triggering inappropriate therapy. An algorithm based upon the confirmation of ventricular activity from a far-field EGM has been developed to reduce inappropriate therapies resulting from this type of lead failure, while ensuring that appropriate therapy is delivered. The objectives of this study were to evaluate the algorithm's ability to discriminate lead noise from ventricular tachycardia/ventricular fibrillation (VT/VF) and to determine whether it inhibits inappropriate shocks without delaying appropriate shocks.
METHODS: The algorithm was prospectively tested using near- and far-field EGM recordings from patients in three conditions: normal sinus rhythm with sustained and non-sustained lead noise via manipulation of the ICD pocket or lead system, and VT/VF induced during defibrillation threshold testing. The recordings were played through a bench-top device running the algorithm with the diagnosis, time to diagnosis, and inhibition of therapy documented.
RESULTS: The algorithm detected noise and withheld inappropriate therapy in 231 of 238 recordings of sustained lead noise that would otherwise have been diagnosed as VT/VF (97.1%). Non-sustained lead noise was correctly diagnosed in 47 of the 52 recordings (90.4%). The device appropriately identified all 853 recordings of VT/VF (100%), without an increase in the time to detection (0.01 ± 0.14 s).
CONCLUSIONS: The SecureSense(TM) algorithm correctly diagnosed sustained and non-sustained lead noise recordings without compromising detection of VT/VF. Use of the algorithm may reduce inappropriate shocks and alert clinicians to lead noise indicative of lead failure.

Entities:  

Keywords:  Defibrillator lead; Implantable cardioverter-defibrillator; Inappropriate therapy; Software

Mesh:

Year:  2016        PMID: 26739482     DOI: 10.1007/s10840-015-0081-6

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  22 in total

1.  Randomized comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from cardiac arrest : the Cardiac Arrest Study Hamburg (CASH).

Authors:  K H Kuck; R Cappato; J Siebels; R Rüppel
Journal:  Circulation       Date:  2000-08-15       Impact factor: 29.690

2.  Single channel ICD noise without loss of conductor circuit integrity.

Authors:  Marc W Deyell; Rutuke K Patel; Joshua M Cooper
Journal:  Pacing Clin Electrophysiol       Date:  2011-10-05       Impact factor: 1.976

3.  Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction.

Authors:  Arthur J Moss; Wojciech Zareba; W Jackson Hall; Helmut Klein; David J Wilber; David S Cannom; James P Daubert; Steven L Higgins; Mary W Brown; Mark L Andrews
Journal:  N Engl J Med       Date:  2002-03-19       Impact factor: 91.245

4.  Amiodarone or an implantable cardioverter-defibrillator for congestive heart failure.

Authors:  Gust H Bardy; Kerry L Lee; Daniel B Mark; Jeanne E Poole; Douglas L Packer; Robin Boineau; Michael Domanski; Charles Troutman; Jill Anderson; George Johnson; Steven E McNulty; Nancy Clapp-Channing; Linda D Davidson-Ray; Elizabeth S Fraulo; Daniel P Fishbein; Richard M Luceri; John H Ip
Journal:  N Engl J Med       Date:  2005-01-20       Impact factor: 91.245

5.  A comparison of antiarrhythmic-drug therapy with implantable defibrillators in patients resuscitated from near-fatal ventricular arrhythmias.

Authors: 
Journal:  N Engl J Med       Date:  1997-11-27       Impact factor: 91.245

6.  Canadian implantable defibrillator study (CIDS) : a randomized trial of the implantable cardioverter defibrillator against amiodarone.

Authors:  S J Connolly; M Gent; R S Roberts; P Dorian; D Roy; R S Sheldon; L B Mitchell; M S Green; G J Klein; B O'Brien
Journal:  Circulation       Date:  2000-03-21       Impact factor: 29.690

7.  Development and testing of an algorithm to detect implantable cardioverter-defibrillator lead failure.

Authors:  Bruce D Gunderson; Jeffrey M Gillberg; Mark A Wood; Pugazhendhi Vijayaraman; Richard K Shepard; Kenneth A Ellenbogen
Journal:  Heart Rhythm       Date:  2006-02       Impact factor: 6.343

8.  Inappropriate implantable cardioverter-defibrillator shocks in MADIT II: frequency, mechanisms, predictors, and survival impact.

Authors:  James P Daubert; Wojciech Zareba; David S Cannom; Scott McNitt; Spencer Z Rosero; Paul Wang; Claudio Schuger; Jonathan S Steinberg; Steven L Higgins; David J Wilber; Helmut Klein; Mark L Andrews; W Jackson Hall; Arthur J Moss
Journal:  J Am Coll Cardiol       Date:  2008-04-08       Impact factor: 24.094

9.  Downloadable algorithm to reduce inappropriate shocks caused by fractures of implantable cardioverter-defibrillator leads.

Authors:  Charles D Swerdlow; Bruce D Gunderson; Kevin T Ousdigian; Athula Abeyratne; Robert W Stadler; Jeffrey M Gillberg; Amisha S Patel; Kenneth A Ellenbogen
Journal:  Circulation       Date:  2008-11-03       Impact factor: 29.690

10.  Prognostic importance of defibrillator shocks in patients with heart failure.

Authors:  Jeanne E Poole; George W Johnson; Anne S Hellkamp; Jill Anderson; David J Callans; Merritt H Raitt; Ramakota K Reddy; Francis E Marchlinski; Raymond Yee; Thomas Guarnieri; Mario Talajic; David J Wilber; Daniel P Fishbein; Douglas L Packer; Daniel B Mark; Kerry L Lee; Gust H Bardy
Journal:  N Engl J Med       Date:  2008-09-04       Impact factor: 91.245

View more
  3 in total

1.  Reduction of inappropriate implantable cardioverter-defibrillator therapies using enhanced supraventricular tachycardia discriminators: the ReduceIT study.

Authors:  Johann Christoph Geller; Armin Wöhrle; Mathias Busch; Albrecht Elsässer; Thomas Kleemann; Frank Birkenhauer; Peter Bramlage; Christian Veltmann
Journal:  J Interv Card Electrophysiol       Date:  2020-07-14       Impact factor: 1.900

2.  Device-Related Complications and Inappropriate Therapies Among Subcutaneous vs. Transvenous Implantable Defibrillator Recipients: Insight Monaldi Rhythm Registry.

Authors:  Vincenzo Russo; Anna Rago; Vincenzo Ruggiero; Francesca Cavaliere; Valter Bianchi; Ernesto Ammendola; Andrea Antonio Papa; Vincenzo Tavoletta; Stefano De Vivo; Paolo Golino; Antonio D'Onofrio; Gerardo Nigro
Journal:  Front Cardiovasc Med       Date:  2022-05-16

3.  Novel ventricular tachyarrhythmia detection enhancement detects undertreated life-threatening arrhythmias.

Authors:  Bruce L Wilkoff; Laurence D Sterns; Michael S Katcher; Gaurav Upadhyay; Peter Seizer; Chaoyi Kang; Jennifer Rhude; Kevin J Davis; Avi Fischer
Journal:  Heart Rhythm O2       Date:  2021-11-18
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.