The impact of technical conditions of X-ray imaging on reproducibility and precision of digital computer-assisted X-ray radiogrammetry (DXR).

A Malich1, J Boettcher, A Pfeil, D Sauner, J P Heyne, A Petrovitch, A Hansch, W Linss, W A Kaiser.   

Abstract

OBJECTIVE: To evaluate the reproducibility of imaging and analysis for bone mineral density (BMD) determination using digital computer-assisted X-ray radiogrammetry (DXR; Pronosco X-posure, version V.2, Sectra Pronosco, Denmark); to verify potential factors that influence BMD extrapolation such as tube voltage, film-focus distance (FFD), film quality and brand (Kodak T-MAT-Plus, Konika SRH, Agfa Scopix), imaging technology (conventional, digital), imaging system (Kodak, Agfa) and exposure level (mAs); and to clarify whether DXR analysis based on printouts of digital images is comparable to analysis of conventional images. DESIGN AND PATIENTS: The hand of a cadaver was X-rayed using varied parameters: 4-8 mAs, 40-52 kV, 90-130 cm FFD. Radiographs under standardised conditions were performed 10 times using a conventional machine (Philips Super 80 CP) and the printouts of a digital system (Digital Diagnost Philips Optimus) for the analysis of reproducibility. One image was scanned and analysed 10 times additionally for imaging reproducibility.
RESULTS: Reliability error of the system for the imaging process using conventional radiographs-rays was 0.49% (standard conditions: 6 mAs, 40 kV, 1 m FFD), using printouts of digital images was 2.89% (4 mAs, 42 kV, 1 m FFD) and regarding the analysis process was 0.22%. BMD calculation is not affected by alterations in FFD (precision error 1.21%), mAs (0.83%) or film quality/brand (0.38%), but differs significantly depending on tube voltage (2.70%). The system was not able to analyse conventional images with tube voltages of 49/52 kV.
CONCLUSION: DXR technology is stable with most of the tested parameters. Normative data should exclusively be used for calculations using similar tube voltage or correction factors. All other parameters had no significant influence on the BMD calculation. Reproducibility is high. For technical reasons it is not recommended to use the printouts of digital images for BMD determination.

Entities:  

Mesh:

Year:  2004        PMID: 15480639     DOI: 10.1007/s00256-004-0861-z

Source DB:  PubMed          Journal:  Skeletal Radiol        ISSN: 0364-2348            Impact factor:   2.199


  13 in total

1.  Reproducibility--an important factor determining the quality of computer-aided detection (CAD) systems.

Authors:  A Malich; T Azhari; T Böhm; M Fleck; W A Kaiser
Journal:  Eur J Radiol       Date:  2000-12       Impact factor: 3.528

2.  Thickness of the cortical layer as an estimate of mineral content of human finger bones.

Authors:  P VIRTAMA; H MAHONEN
Journal:  Br J Radiol       Date:  1960-01       Impact factor: 3.039

3.  The radiological diagnosis of osteoporosis: a new approach.

Authors:  E BARNETT; B E NORDIN
Journal:  Clin Radiol       Date:  1960-07       Impact factor: 2.350

4.  Effects of growth hormone treatment in children with cystic fibrosis: the National Cooperative Growth Study experience.

Authors:  D S Hardin; J P Sy
Journal:  J Pediatr       Date:  1997-07       Impact factor: 4.406

5.  Recent developments in DXA. Quality of new DXA/MXA-devices for densitometry and morphometry.

Authors:  D Felsenberg; W Gowin; E Diessel; S Armbrust; J Mews
Journal:  Eur J Radiol       Date:  1995-09       Impact factor: 3.528

6.  The accuracy of peripheral skeletal assessment at the radius in estimating femoral bone density as measured by dual-energy X-ray absorptiometry: a comparative study of single-photon absorptiometry and computed tomography.

Authors:  S Boonen; X Cheng; P H Nicholson; G Verbeke; P Broos; J Dequeker
Journal:  J Intern Med       Date:  1997-10       Impact factor: 8.989

7.  Ultrasound bone densitometry of the os calcis in children and adolescents.

Authors:  K Wünsche; B Wünsche; H Fähnrich; H J Mentzel; S Vogt; K Abendroth; W A Kaiser
Journal:  Calcif Tissue Int       Date:  2000-11       Impact factor: 4.333

Review 8.  Radiation exposure in bone mineral density assessment.

Authors:  C F Njeh; T Fuerst; D Hans; G M Blake; H K Genant
Journal:  Appl Radiat Isot       Date:  1999-01       Impact factor: 1.513

9.  Measurement of bone mineral content of the lumbar spine by dual energy x-ray absorptiometry in normal children: correlations with growth parameters.

Authors:  C Glastre; P Braillon; L David; P Cochat; P J Meunier; P D Delmas
Journal:  J Clin Endocrinol Metab       Date:  1990-05       Impact factor: 5.958

10.  Normative values of bone parameters of children and adolescents using digital computer-assisted radiogrammetry (DXR).

Authors:  Ansgar Malich; Martin G Freesmeyer; Hans-Joachim Mentzel; Dieter Sauner; Joachim Boettcher; Alexander Petrovitch; Werner Behrendt; Werner A Kaiser
Journal:  J Clin Densitom       Date:  2003       Impact factor: 2.963

View more
  7 in total

1.  Feasibility study of semi-automated measurements of finger joint space widths.

Authors:  Alexander Pfeil; Julia Sommerfeld; Rosemarie Fröber; Gabriele Lehmann; Ansgar Malich; Andreas Hansch; Gunter Wolf; Joachim Böttcher
Journal:  Rheumatol Int       Date:  2010-04-17       Impact factor: 2.631

Review 2.  CAD for mammography: the technique, results, current role and further developments.

Authors:  Ansgar Malich; Dorothee R Fischer; Joachim Böttcher
Journal:  Eur Radiol       Date:  2006-01-17       Impact factor: 5.315

3.  Long-term in-vitro precision of direct digital X-ray radiogrammetry.

Authors:  Alvilde Dhainaut; Mari Hoff; Johan Kälvesten; Stian Lydersen; Kristina Forslind; Glenn Haugeberg
Journal:  Skeletal Radiol       Date:  2011-05-11       Impact factor: 2.199

4.  Evaluation of bone-mineral density by digital X-ray radiogrammetry (DXR) in pediatric renal transplant recipients.

Authors:  Hans-J Mentzel; Ulrike John; Joachim Boettcher; Ansgar Malich; Alexander Pfeil; Rüdiger Vollandt; Joachim Misselwitz; Werner A Kaiser
Journal:  Pediatr Radiol       Date:  2004-12-29

5.  Digital X-ray radiogrammetry and its sensitivity and specificity for the identification of rheumatoid arthritis-related cortical hand bone loss.

Authors:  Alexander Pfeil; Glenn Haugeberg; Diane M Renz; Lisa Reinhardt; Christian Jung; Marcus Franz; Gunter Wolf; Joachim Böttcher
Journal:  J Bone Miner Metab       Date:  2016-03-15       Impact factor: 2.626

6.  Early changes in bone mineral density measured by digital X-ray radiogrammetry predict up to 20 years radiological outcome in rheumatoid arthritis.

Authors:  Meliha C Kapetanovic; Elisabet Lindqvist; Jakob Algulin; Kjell Jonsson; Tore Saxne; Kerstin Eberhardt; Pierre Geborek
Journal:  Arthritis Res Ther       Date:  2011-02-23       Impact factor: 5.156

7.  Metacarpal Index Estimated by Digital X-ray Radiogrammetry as a Tool for Differentiating Rheumatoid Arthritis Related Periarticular Osteopenia.

Authors:  Joachim Böttcher; Alexander Pfeil; Alexander Petrovitch; Mirco Schmidt; Anika Kramer; Max Ludwig Schäfer; Mieczyslaw Gajda; Gert Hein; Gunter Wolf; Werner A Kaiser
Journal:  Int J Biomed Sci       Date:  2006-09
  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.