Doppler assessment of mechanical aortic valve prostheses: effect of valve design and size of the aorta.

Julia Mascherbauer1, Heinrich Schima, Gerald Maurer, Helmut Baumgartner.   

Abstract

BACKGROUND AND AIM OF THE STUDY: Discrepancies between Doppler and catheter gradients have been reported for bileaflet aortic valve prostheses. Whether modifications in geometric design of newly developed bileaflet valves lead to a different Doppler-catheter gradient relationship has not been evaluated. Variable results have been reported for tilting-disc prostheses. In addition, the effect of aortic size on the Doppler-catheter gradient relationship remains unclear.
METHODS: Various sizes of On-X and Edwards Mira (identical with Sorin Bicarbon) bileaflet valves and Sorin Allcarbon tilting-disc aortic valves (19-25 mm) were studied in a pulsatile flow model. Doppler and catheter gradients were measured simultaneously. Aortic diameters between 1.8 and 4 cm were evaluated.
RESULTS: Correlation between Doppler and catheter gradients was excellent (r = 0.98-0.99 for peak and mean gradients), but in bileaflet valves Doppler significantly overestimated the corresponding catheter gradients as reflected by slopes of the regression lines (1.57-1.8). In the range of relevant gradients > or = 10 mmHg, Doppler exceeded catheter gradients by 40 +/- 17% (peak) and 39 +/- 16% (mean) in Mira valves, and by 46 +/- 19% (peak) and 43 +/- 14% (mean) in On-X valves. In the Sorin tilting-disc valve, Doppler accurately reflected catheter gradients (slopes of regression lines 1.05-1.14). The aortic diameter significantly influenced results in only tilting-disc valves, but in absolute terms the effect was clinically less relevant.
CONCLUSION: Discrepancies between Doppler and catheter gradients are common to all bileaflet valves, regardless of their specific geometric design, whereas tilting-disc valves must be considered individually. The influence of aortic size on the Doppler-catheter gradient relationship appears clinically to be less relevant in prosthetic valves.

Mesh:

Year:  2004        PMID: 15473486

Source DB:  PubMed          Journal:  J Heart Valve Dis        ISSN: 0966-8519


  2 in total

1.  Clinical and echocardiographic assessment of the Medtronic Advantage aortic valve prosthesis: the Scandinavian multicentre, prospective study.

Authors:  Rune Haaverstad; Nicola Vitale; Asbjørn Karevold; Giangiuseppe Cappabianca; Arve Tromsdal; Peter Skov Olsen; Lars Köber; Halfdan Ihlen; Kjell Arne Rein; Jan L Svennevig
Journal:  Heart       Date:  2006-10-25       Impact factor: 5.994

2.  Comparison of Transvalvular Aortic Mean Gradients Obtained by Intraprocedural Echocardiography and Invasive Measurement in Balloon and Self-Expanding Transcatheter Valves.

Authors:  Amr E Abbas; Ramy Mando; Amer Kadri; Houman Khalili; George Hanzel; Francis Shannon; Karim Al-Azizi; Thomas Waggoner; Safwan Kassas; Thomas Pilgrim; Taishi Okuno; Alexander Camacho; Alexandra Selberg; Sammy Elmariah; Anthony Bavry; Julien Ternacle; Jared Christensen; Neil Gheewala; Philippe Pibarot; Michael Mack
Journal:  J Am Heart Assoc       Date:  2021-09-29       Impact factor: 5.501

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.