Accuracy of automated auscultatory blood pressure measurement during supine exercise and treadmill stress electrocardiogram-testing.

James D Cameron1, Irene Stevenson, Emily Reed, Barry P McGrath, Anthony M Dart, Bronwyn A Kingwell.   

Abstract

OBJECTIVES: Monitoring of brachial blood pressure during exercise-electrocardiogram (ECG) testing is mandatory and changes in blood pressure (BP) can provide critical management evidence. Patient movement, mechanical vibration, artifactual sounds and observer variability make standard manual techniques problematic. This was an investigator-initiated study to assess an automated auscultatory technique of BP assessment [Tango exercise blood pressure monitor (SunTech Medical Instruments, NC, USA)] to adequately measure BP during stress-ECG testing.
METHODS: Initially five fit young male volunteers underwent invasive right brachial artery BP recording using a low-compliance fluid-filled catheter with simultaneous manual and automated assessment. Secondarily, during exercise-ECG testing, the system was assessed against beat-to-beat brachial blood pressures obtained from a catheter-tip solid-state pressure manometer positioned in the ipsilateral brachial artery.
RESULTS: In the supine study overall mean difference (+/-SEM) between invasive and manual blood pressures was 3.26 (1.53) and 3.89 (1.90) mmHg for diastolic BP (DBP) and systolic BP (SBP) respectively. Corresponding differences between invasive and automated results, and manual and automated were 3.68 (0.84) and -7.31 (1.83) mmHg, and -0.64 (+/-1.43) and -11.42 (+/-1.59) mmHg. During treadmill exercise-ECG testing the combined mean difference (+/-SEM) between invasive and automated SBP and DBP was 4.79 (+/-0.14) and 6.33 (+/-0.10) mmHg, respectively.
CONCLUSION: Automated BP assessment during exercise-ECG testing is feasible with the use of appropriate automatic devices likely to be at least as accurate as manual BP registration. The Tango device is tolerant to exercise and provides reliable automatic BP assessment with absolute differences within an acceptable clinical range.

Entities:  

Mesh:

Year:  2004        PMID: 15472500     DOI: 10.1097/00126097-200410000-00007

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  20 in total

1.  Effect of muscle metaboreflex activation on spontaneous cardiac baroreflex sensitivity during exercise in humans.

Authors:  Doreen Hartwich; William E Dear; Jessica L Waterfall; James P Fisher
Journal:  J Physiol       Date:  2011-10-03       Impact factor: 5.182

2.  The influence of aerobic exercise training on the double product break point in low-to-moderate risk adults.

Authors:  Trent A Hargens; Diane C Griffin; Leonard A Kaminsky; Mitchell H Whaley
Journal:  Eur J Appl Physiol       Date:  2010-09-28       Impact factor: 3.078

3.  Exercise testing is useful to screen for residual coarctation in children.

Authors:  Bibhuti B Das; Shashi Raj; Lawrence Shoemaker
Journal:  Pediatr Cardiol       Date:  2009-05-16       Impact factor: 1.655

4.  Exercise Blood Pressure Guidelines: Time to Re-evaluate What is Normal and Exaggerated?

Authors:  Katharine D Currie; John S Floras; Andre La Gerche; Jack M Goodman
Journal:  Sports Med       Date:  2018-08       Impact factor: 11.136

5.  Average daily blood pressure, not office blood pressure, is associated with progression of cerebrovascular disease and cognitive decline in older people.

Authors:  William B White; Leslie Wolfson; Dorothy B Wakefield; Charles B Hall; Patrick Campbell; Nicola Moscufo; Julia Schmidt; Richard F Kaplan; Godfrey Pearlson; Charles R G Guttmann
Journal:  Circulation       Date:  2011-11-22       Impact factor: 29.690

6.  Blood pressure and leg deoxygenation are exaggerated during treadmill walking in patients with peripheral artery disease.

Authors:  Amanda J Miller; J Carter Luck; Danielle Jin-Kwang Kim; Urs A Leuenberger; David N Proctor; Lawrence I Sinoway; Matthew D Muller
Journal:  J Appl Physiol (1985)       Date:  2017-08-17

Review 7.  Exercise blood pressure: clinical relevance and correct measurement.

Authors:  J E Sharman; A LaGerche
Journal:  J Hum Hypertens       Date:  2014-10-02       Impact factor: 3.012

8.  Cardiopulmonary exercise testing in congenital heart disease: equipment and test protocols.

Authors:  T Takken; A C Blank; E H Hulzebos; M van Brussel; W G Groen; P J Helders
Journal:  Neth Heart J       Date:  2009-09       Impact factor: 2.380

9.  Spontaneous baroreflex measures are unable to detect age-related impairments in cardiac baroreflex function during dynamic exercise in humans.

Authors:  James P Fisher; Shigehiko Ogoh; Chelif Junor; Azamuddin Khaja; Mindy Northrup; Paul J Fadel
Journal:  Exp Physiol       Date:  2009-01-12       Impact factor: 2.969

10.  Association of von Willebrand factor blood levels with exercise hypertension.

Authors:  Sonja B Nikolic; Murray J Adams; Petr Otahal; Lindsay M Edwards; James E Sharman
Journal:  Eur J Appl Physiol       Date:  2014-12-25       Impact factor: 3.078

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.