| Literature DB >> 15465853 |
Harminder Bajaj1, Vikas K Sharma, Devendra S Kalonia.
Abstract
A method is proposed for the measurement of the B22 value of proteins in aqueous solutions in flow-mode that utilizes a novel fabricated dual-detector cell, which simultaneously measures protein concentration and the corresponding scattered light intensity at 90 degrees , after the protein elutes from a size-exclusion column. Each data point on the chromatograms obtained from the light scattering detector and the concentration (ultraviolet) detector is converted to Rayleigh's ratio, Rtheta, and concentration, c, respectively. The B22 value is calculated from the slope of the Debye plot (Kc/Rtheta versus c) generated from a range of concentrations obtained from these chromatograms for a single protein injection. It is shown that this method provides reliable determination of the B22 values for such proteins as lysozyme, chymotrypsinogen, and chymotrypsin in various solution conditions that agree well with those reported in literature.Mesh:
Substances:
Year: 2004 PMID: 15465853 PMCID: PMC1304913 DOI: 10.1529/biophysj.104.048686
Source DB: PubMed Journal: Biophys J ISSN: 0006-3495 Impact factor: 4.033