BACKGROUND: In 2003, the Centers for Medicaid and Medicare Services recommended QRS duration as a means to identify MADIT II-like patients suitable for implanted cardiac defibrillator (ICD) therapy. We compared the ability of microvolt T-wave alternans and QRS duration to identify groups at high and low risk of dying among heart failure patients who met MADIT II criteria for ICD prophylaxis. METHODS AND RESULTS: Patients with MADIT II characteristics and sinus rhythm had a microvolt T-wave alternans exercise test and a 12-lead ECG. Our primary end point was 2-year all-cause mortality. Of 177 MADIT II-like patients, 32% had a QRS duration >120 ms, and 68% had an abnormal (positive or indeterminate) microvolt T-wave alternans test. During an average follow-up of 20+/-6 months, 20 patients died. We compared patients with an abnormal microvolt T-wave alternans test to those with a normal (negative) test, and patients with a QRS >120 ms with those with a QRS < or =120 ms; the hazard ratios for 2-year mortality were 4.8 (P=0.020) and 1.5 (P=0.367), respectively. The actuarial mortality rate was substantially lower among patients with a normal microvolt T-wave alternans test (3.8%; 95% confidence interval: 0, 9.0) than the mortality rate in patients with a narrow QRS (12.0%; 95% confidence interval: 5.6, 18.5). The corresponding false-negative rates are 3.5% and 10.2%, respectively. CONCLUSIONS: Among MADIT II-like patients, a microvolt T-wave alternans test is better than QRS duration at identifying a high-risk group and also better at identifying a low-risk group unlikely to benefit from ICD therapy.
BACKGROUND: In 2003, the Centers for Medicaid and Medicare Services recommended QRS duration as a means to identify MADIT II-like patients suitable for implanted cardiac defibrillator (ICD) therapy. We compared the ability of microvolt T-wave alternans and QRS duration to identify groups at high and low risk of dying among heart failurepatients who met MADIT II criteria for ICD prophylaxis. METHODS AND RESULTS:Patients with MADIT II characteristics and sinus rhythm had a microvolt T-wave alternans exercise test and a 12-lead ECG. Our primary end point was 2-year all-cause mortality. Of 177 MADIT II-like patients, 32% had a QRS duration >120 ms, and 68% had an abnormal (positive or indeterminate) microvolt T-wave alternans test. During an average follow-up of 20+/-6 months, 20 patients died. We compared patients with an abnormal microvolt T-wave alternans test to those with a normal (negative) test, and patients with a QRS >120 ms with those with a QRS < or =120 ms; the hazard ratios for 2-year mortality were 4.8 (P=0.020) and 1.5 (P=0.367), respectively. The actuarial mortality rate was substantially lower among patients with a normal microvolt T-wave alternans test (3.8%; 95% confidence interval: 0, 9.0) than the mortality rate in patients with a narrow QRS (12.0%; 95% confidence interval: 5.6, 18.5). The corresponding false-negative rates are 3.5% and 10.2%, respectively. CONCLUSIONS: Among MADIT II-like patients, a microvolt T-wave alternans test is better than QRS duration at identifying a high-risk group and also better at identifying a low-risk group unlikely to benefit from ICD therapy.
Authors: Euler de Vilhena Garcia; Nelson Samesima; Horácio G Pereira Filho; Cristina M Quadros; Luis Tenório Cavalcante da Silva; Martino Martinelli Filho; Maria Luciana Zacharias Hannouche; Wilson Mathias; Carlos Alberto Pastore Journal: Ann Noninvasive Electrocardiol Date: 2009-04 Impact factor: 1.468
Authors: Michael R Gold; John H Ip; Otto Costantini; Jeanne E Poole; Steven McNulty; Daniel B Mark; Kerry L Lee; Gust H Bardy Journal: Circulation Date: 2008-10-27 Impact factor: 29.690
Authors: Lance D Wilson; Darwin Jeyaraj; Xiaoping Wan; Gregory S Hoeker; Tamer H Said; Matthew Gittinger; Kenneth R Laurita; David S Rosenbaum Journal: Heart Rhythm Date: 2008-11-08 Impact factor: 6.343