Literature DB >> 15446521

Corrosion of cemented titanium femoral stems.

S R Thomas1, D Shukla, P D Latham.   

Abstract

Cemented titanium stems in hip arthroplasty are associated with proximal cement-stem ebonding and early failure. This was well publicised with the 3M Capital hip. However, corrosion in this setting has been reported with only one stem design and is less widely accepted. We present a series of 12 cemented titanium Furlong Straight Stems which required revision at a mean of 78 months for thigh pain. At revision the stems were severely corroded in a pattern which was typical of crevice corrosion. Symptoms were eliminated after revision to an all-stainless steel femoral prosthesis of the same design. We discuss the likely causes for the corrosion. The combination of a titanium stem and cement appears to facilitate crevice corrosion.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15446521     DOI: 10.1302/0301-620x.86b7.14812

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  7 in total

1.  Clinical and biological assessment of cemented titanium femoral stems: an 11-year experience.

Authors:  Patrick Boyer; Jean-Yves Lazennec; Joel Poupon; Marc-Antoine Rousseau; Philippe Ravaud; Yves Catonné
Journal:  Int Orthop       Date:  2008-10-28       Impact factor: 3.075

2.  Clinical significance of corrosion of cemented femoral stems in metal-on-metal hips: a retrieval study.

Authors:  Harry S Hothi; Reshid Berber; Andreas C Panagiotopoulos; Robert K Whittaker; Camilla Rhead; John A Skinner; Alister J Hart
Journal:  Int Orthop       Date:  2016-01-25       Impact factor: 3.075

3.  Lipopolysaccharide inhibits or accelerates biomedical titanium corrosion depending on environmental acidity.

Authors:  Fei Yu; Owen Addison; Stephen J Baker; Alison J Davenport
Journal:  Int J Oral Sci       Date:  2015-09-14       Impact factor: 6.344

4.  Cytokine Profile in Patients with Aseptic Loosening of Total Hip Replacements and Its Relation to Metal Release and Metal Allergy.

Authors:  Rune J Christiansen; Henrik J Münch; Charlotte M Bonefeld; Jacob P Thyssen; Jens J Sloth; Carsten Geisler; Kjeld Søballe; Morten S Jellesen; Stig S Jakobsen
Journal:  J Clin Med       Date:  2019-08-20       Impact factor: 4.241

5.  Corrosion of Polished Cobalt-Chrome Stems Presenting as Cobalt Encephalopathy.

Authors:  Stephen S Tower; Dana J Medlin; Robert L Bridges; Christina S Cho
Journal:  Arthroplast Today       Date:  2020-12-02

6.  18 years of results with cemented primary hip prostheses in the Norwegian Arthroplasty Register: concerns about some newer implants.

Authors:  Birgitte Espehaug; Ove Furnes; Lars B Engesaeter; Leif I Havelin
Journal:  Acta Orthop       Date:  2009-08       Impact factor: 3.717

7.  10-year results of a new low-monomer cement: follow-up of a randomized RSA study.

Authors:  Per Söderlund; Jon Dahl; Stephan Röhrl; Bo Nivbrant; Kjell G Nilsson
Journal:  Acta Orthop       Date:  2012-11-01       Impact factor: 3.717

  7 in total