Perfusion technique for nonhaemic cardiopulmonary bypass prime in neonates and infants under 6 kg body weight.

Frank Merkle1, Wolfgang Boettcher, Fritz Schulz, Andreas Koster, Michael Huebler, Roland Hetzer.   

Abstract

BACKGROUND: Cardiopulmonary bypass (CPB) in neonates and infants is associated with significant haemodilution when priming of the CPB circuit is accomplished without transfusion of homologous blood components. The degree of haemodilution and, thus, the requirements for blood transfusion may be reduced when the CPB circuit is miniaturized without compromising patient safety.
METHOD: Between January 2002 and October 2003, selected neonates and small infants were operated on using a nonhaemic prime extracorporeal circuit. CPB priming volume could be reduced from 300 mL to 190 mL by using a dedicated neonatal CPB console with mast-mounted roller pump heads. Reduction of priming volume resulted from shortening of all CPB lines to the minimum, downsizing of all CPB lines, exclusion of unused CPB components, use of vacuum-assisted venous drainage and from close co-operation between the perfusionist, cardiac surgeon and anaesthesiologist. The reduction in priming volume was achieved without eliminating the arterial line filter as safety device.
RESULTS: A total of nine patients weighing between 3.2 and 5.9 kg (mean 4.7 kg) and with a body surface area of 0.22-0.35 m2 (mean 0.29 m2) were operated on with the use of the modified neonatal CPB circuit and a nonhaemic prime. Bypass time varied from 38 to 167 min (mean 96 min). The mean haematocrit on CPB was 22.5% with a range of 17-29%. The postoperative course of all patients was uneventful.
CONCLUSION: A significant reduction in CPB priming volume makes nonhaemic prime CPB in neonates and small infants undergoing complex repair of congenital heart defects possible.

Entities:  

Mesh:

Year:  2004        PMID: 15376767     DOI: 10.1191/0267659104pf744oa

Source DB:  PubMed          Journal:  Perfusion        ISSN: 0267-6591            Impact factor:   1.972


  6 in total

1.  The neonatal circuit: in search of the ultimate solution.

Authors:  Filip De Somer
Journal:  J Extra Corpor Technol       Date:  2012-03

Review 2.  Vacuum-assisted venous drainage and gaseous microemboli in cardiopulmonary bypass.

Authors:  Shigang Wang; Akif Undar
Journal:  J Extra Corpor Technol       Date:  2008-12

3.  Vacuum assist: angel or demon CON.

Authors:  Timothy W Willcox
Journal:  J Extra Corpor Technol       Date:  2013-06

4.  Bloodless Repair for a 3.6 Kilogram Transposition of the Great Arteries with Jehovah's Witness Faith.

Authors:  Jeffery L Burnside; Todd M Ratliff; Ashley B Hodge; Daniel Gomez; Mark Galantowicz; Aymen Naguib
Journal:  J Extra Corpor Technol       Date:  2017-12

5.  The Impact of Roller Pump vs. Centrifugal Pump on Homologous Blood Transfusion in Pediatric Cardiac Surgery.

Authors:  Bharat Datt; Moui B Nguyen; Gary Plancher; Mark Ruzmetov; Michael O'Brien; Alicia Kube; Hamish M Munro; Kamal K Pourmoghadam; William M DeCampli
Journal:  J Extra Corpor Technol       Date:  2017-03

6.  Transfusion-free cardiopulmonary bypass in Jehovah's Witness patients weighing less than 5 kg.

Authors:  Wolfgang Boettcher; Frank Merkle; Michael Huebler; Andreas Koster; Fritz Schulz; Michael Kopitz; Hermann Kuppe; Peter Lange; Roland Hetzer
Journal:  J Extra Corpor Technol       Date:  2005-09
  6 in total