Measurement of meniscofemoral contact pressure after repair of bucket-handle tears with biodegradable implants.

Roland Becker1, Dieter Wirz, Cornelius Wolf, Beat Göpfert, Wolfgang Nebelung, Niklaus Friederich.   

Abstract

INTRODUCTION: Biodegradable implants are frequently used for meniscus repair. Articular cartilage damage has been reported recently after meniscus repair with biodegradable implants. The aim of the study was to investigate the meniscofemoral contact pressure at the posterior horn of the medial and lateral meniscus after repair of bucket-handle lacerations.
MATERIALS AND METHODS: Specimens were mounted in a materials testing machine (Bionix 858, MTS) which was equipped with a load cell. The quadriceps tendon was attached to a hydraulic cylinder, and knee motion was controlled via tension of the quadriceps tendon. A piezo-resistive system (Tekscan, Boston, MA, USA) measured the meniscofemoral contact pressure. Five different types of biodegradable implants (Arrow, Dart, Fastener, Stinger and Meniscal Screw) and horizontal suture (no. 2 Ethibond) were tested. The knee was extended from 90 degrees of flexion to 0 degrees under a constant load of 350 N due to adjustment of the tension force of the quadriceps tendon. The femorotibial pressure and contact area were recorded at 0 degree, 30 degrees, 60 degrees and 90 degrees of flexion.
RESULTS: The meniscofemoral pressure did not increase after meniscus repair with biodegradable implants or sutures. The meniscofemoral peak pressure at the posterior horn was 1.46+/-1.54 MPa in the medial compartment and 1.08+/-1.17 MPa in the lateral compartment at full knee extension. The meniscofemoral pressure increased significantly in both compartments with knee flexion from 0 degree to 90 degrees.
CONCLUSION: Biodegradable implants for meniscus repair do not affect the meniscofemoral pressure. However, there remains a risk of damage to the cartilage when barbed implants are used. If the implant is not entirely advanced into the meniscus, the sharp head or some of the barbs at the column of the implant may come into direct contact with the articular cartilage of the femoral condyle or tibial plateau. The authors presume that incorrect positioning of the implant seems to be the major reason for cartilage damage.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15365717     DOI: 10.1007/s00402-004-0739-5

Source DB:  PubMed          Journal:  Arch Orthop Trauma Surg        ISSN: 0936-8051            Impact factor:   3.067


  17 in total

1.  Rehabilitation following meniscal repair.

Authors:  John T Cavanaugh; Sarah E Killian
Journal:  Curr Rev Musculoskelet Med       Date:  2012-03

2.  Effect of media mixing on ECM assembly and mechanical properties of anatomically-shaped tissue engineered meniscus.

Authors:  Jeffrey J Ballyns; Timothy M Wright; Lawrence J Bonassar
Journal:  Biomaterials       Date:  2010-06-12       Impact factor: 12.479

3.  Medial meniscus anatomy-from basic science to treatment.

Authors:  Robert Śmigielski; Roland Becker; Urszula Zdanowicz; Bogdan Ciszek
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-12-24       Impact factor: 4.342

Review 4.  Joint contact stress: a reasonable surrogate for biological processes?

Authors:  Richard A Brand
Journal:  Iowa Orthop J       Date:  2005

Review 5.  [Ramp lesions : Tips and tricks in diagnostics and therapy].

Authors:  R Seil; A Hoffmann; S Scheffler; D Theisen; C Mouton; D Pape
Journal:  Orthopade       Date:  2017-10       Impact factor: 1.087

6.  Contact mechanics after mattress suture repair of medial meniscus vertical longitudinal tear: an in vitro study.

Authors:  Zhong Chen; Haozhi Zhang; Huan Luo; Rui Yang; Zhengzheng Zhang; Chuan Jiang; Jingyi Hou; Yunfeng Zhou; Yue Xu; Bin Song; Weiping Li
Journal:  Arch Orthop Trauma Surg       Date:  2020-04-18       Impact factor: 3.067

7.  Arthroscopic Saucerization of a Symptomatic Posterior Horn Tear in a Discoid Medial Meniscus.

Authors:  Bhumit Desai; Michael Warren; Lacey G Lavie; Michael Nammour; Sean Waldron
Journal:  Ochsner J       Date:  2021

8.  A pre-clinical test platform for the functional evaluation of scaffolds for musculoskeletal defects: the meniscus.

Authors:  Suzanne A Maher; Scott A Rodeo; Hollis G Potter; Lawrence J Bonassar; Timothy M Wright; Russell F Warren
Journal:  HSS J       Date:  2011-01-08

9.  Validation of a method for combining biplanar radiography and magnetic resonance imaging to estimate knee cartilage contact.

Authors:  Eric Thorhauer; Scott Tashman
Journal:  Med Eng Phys       Date:  2015-08-21       Impact factor: 2.242

10.  PLDLA/PCL-T Scaffold for Meniscus Tissue Engineering.

Authors:  Andrea Rodrigues Esposito; Marlon Moda; Silvia Mara de Melo Cattani; Gracy Mara de Santana; Juliana Abreu Barbieri; Monique Moron Munhoz; Túlio Pereira Cardoso; Maria Lourdes Peris Barbo; Teresa Russo; Ugo D'Amora; Antonio Gloria; Luigi Ambrosio; Eliana Aparecida de Rezende Duek
Journal:  Biores Open Access       Date:  2013-04
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.