Vascular complications associated with arteriotomy closure devices in patients undergoing percutaneous coronary procedures: a meta-analysis.

Eugenia Nikolsky1, Roxana Mehran, Amir Halkin, Eve D Aymong, Gary S Mintz, Zoran Lasic, Manuela Negoita, Martin Fahy, Shoshana Krieger, Issam Moussa, Jeffrey W Moses, Gregg W Stone, Martin B Leon, Stuart J Pocock, George Dangas.   

Abstract

OBJECTIVES: This study was designed to assess the safety of arteriotomy closure devices (ACDs) versus mechanical compression by meta-analysis in patients undergoing percutaneous transfemoral coronary procedures.
BACKGROUND: Although ACDs are widely applied for hemostasis after percutaneous endovascular procedures, their safety is controversial.
METHODS: Randomized, case-control, and cohort studies comparing access-related complications using ACDs versus mechanical compression were analyzed. The primary end point was the cumulative incidence of vascular complications, including pseudoaneurysm, arteriovenous fistula, retroperitoneal hematoma, femoral artery thrombosis, surgical vascular repair, access site infection, and blood transfusion.
RESULTS: A total of 30 studies involving 37,066 patients were identified. No difference in complication incidence between Angio-Seal and mechanical compression was revealed in the diagnostic (Dx) setting (odds ratio [OR] 1.08, 95% confidence interval [CI] 0.11 to 10.0) or percutaneous coronary interventions (PCI) (OR 0.86, 95% CI 0.65 to 1.12). Meta-analysis of randomized trials only showed a trend toward less complications using Angio-Seal in a PCI setting (OR 0.46, 95% CI 0.20 to 1.04; p = 0.062). No differences were observed regarding Perclose in either Dx (OR 1.51, 95% CI 0.24 to 9.47) or PCI (OR 1.21, 95% CI 0.94 to 1.54) setting. An increased risk in complication rates using VasoSeal in the PCI setting (OR 2.25, 95% CI 1.07 to 4.71) was found. The overall analysis favored mechanical compression over ACD (OR 1.34, 95% CI 1.01 to 1.79).
CONCLUSIONS: In the setting of Dx angiography, the risk of access-site-related complications was similar for ACD compared with mechanical compression. In the setting of PCI, the rate of complications appeared higher with VasoSeal.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15364320     DOI: 10.1016/j.jacc.2004.06.048

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  61 in total

1.  Embolisation of Angio-Seal™ device: an unusual case of post-cardiac catheterisation limb ischaemia.

Authors:  R S Khakha; T Ali; P Chong; D Gerrard; P Leopold
Journal:  Ann R Coll Surg Engl       Date:  2011-07       Impact factor: 1.891

2.  Radioembolization complicated by dissection of the common femoral artery.

Authors:  Sarah G Deitch; Ramona Gupta
Journal:  Semin Intervent Radiol       Date:  2011-06       Impact factor: 1.513

Review 3.  Radial versus femoral access for percutaneous coronary intervention: implications for vascular complications and bleeding.

Authors:  Sandeep Nathan; Sunil V Rao
Journal:  Curr Cardiol Rep       Date:  2012-08       Impact factor: 2.931

4.  Safety and feasibility of a novel vascular closure device in neurointerventional procedures.

Authors:  Ammar Taha; Ellen K Walsh; Kathryn A Wright; Iftikhar Ahmed; Nucharin Supakul; Eric E Awwad; Juan G Tejada
Journal:  Interv Neuroradiol       Date:  2013-09-26       Impact factor: 1.610

5.  Critical lower limb ischemia from an embolized Angio-Seal closure device.

Authors:  Chris Cianci; Robert C Kowal; Georges Feghali; Stephen Hohmann; Robert C Stoler; James W Choi
Journal:  Proc (Bayl Univ Med Cent)       Date:  2013-10

6.  Recommendations for Endovascular Care of Stroke Patients.

Authors:  Michelle Hill; Brenda A Glenn; Brenda J Reese; Benjamin Morrow
Journal:  Interv Neurol       Date:  2017-11-17

7.  2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias.

Authors:  Edmond M Cronin; Frank M Bogun; Philippe Maury; Petr Peichl; Minglong Chen; Narayanan Namboodiri; Luis Aguinaga; Luiz Roberto Leite; Sana M Al-Khatib; Elad Anter; Antonio Berruezo; David J Callans; Mina K Chung; Phillip Cuculich; Andre d'Avila; Barbara J Deal; Paolo Della Bella; Thomas Deneke; Timm-Michael Dickfeld; Claudio Hadid; Haris M Haqqani; G Neal Kay; Rakesh Latchamsetty; Francis Marchlinski; John M Miller; Akihiko Nogami; Akash R Patel; Rajeev Kumar Pathak; Luis C Saenz Morales; Pasquale Santangeli; John L Sapp; Andrea Sarkozy; Kyoko Soejima; William G Stevenson; Usha B Tedrow; Wendy S Tzou; Niraj Varma; Katja Zeppenfeld
Journal:  J Interv Card Electrophysiol       Date:  2020-10       Impact factor: 1.900

8.  The data extraction and longitudinal trend analysis network study of distributed automated postmarket cardiovascular device safety surveillance.

Authors:  Amit Kumar; Michael E Matheny; Kalon K L Ho; Robert W Yeh; Thomas C Piemonte; Howard Waldman; Pinak B Shah; Richard Cope; Sharon-Lise T Normand; Sharon Donnelly; Susan Robbins; Frederic S Resnic
Journal:  Circ Cardiovasc Qual Outcomes       Date:  2014-12-09

9.  Late onset iatrogenic limb ischaemia after deployment of an Angio-Seal vascular closure device.

Authors:  Rama K Krishna; Nisharahmed Kherada; Nirat Beohar
Journal:  BMJ Case Rep       Date:  2015-04-26

10.  ARFI ultrasound for in vivo hemostasis assessment postcardiac catheterization, part II: pilot clinical results.

Authors:  Russell H Behler; Mallory R Scola; Timothy C Nichols; Melissa C Caughey; Melrose W Fisher; Hongtu Zhu; Caterina M Gallippi
Journal:  Ultrason Imaging       Date:  2009-07       Impact factor: 1.578

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.