Aortoiliac gadolinium-enhanced CT angiography: improved results with a 16-detector row scanner compared with a four-detector row scanner.

Stephan Wicky1, Alan Greenfield, Chieh-Min Fan, Stuart C Geller, Leena M Hamberg, Udo Hoffmann, Arthur C Waltman.   

Abstract

PURPOSE: To assess the level of vascular enhancement of gadolinium-enhanced aortoiliac computed tomographic (CT) angiography with a 16-detector row CT scanner and to compare it with the results of previous similar studies that used four-detector row CT units.
MATERIALS AND METHODS: Eleven gadolinium-enhanced CT angiograms were obtained in 10 consecutive patients with contraindication to iodinated contrast medium with use of a 16-detector row CT scanner. In the region of interest, attenuation measurements (in HU) were obtained from the proximal abdominal aorta to the common femoral arteries during unenhanced, gadolinium-enhanced, and delayed acquisitions. The results were compared to those in the 15 consecutive patients who most recently had similar examinations performed on a four-detector row CT unit. Phantom studies with diluted gadolinium were conducted to compare attenuation between CT units.
RESULTS: On four-detector row CT, throughout the scan length, mean enhancement values were 53.8 HU +/- 5.3 and 15.0 HU +/- 2.6 for gadolinium-enhanced and delayed series, respectively. For the 16-detector row CT unit, they were 76.1 HU +/- 3.4 and 21.3 HU +/- 1.3, respectively. As a result of a shorter scan time and a more optimal start time, the 16-detector row CT unit provided significantly greater and more consistent enhancement throughout the scan length compared with the four-detector row CT unit (P =.0106). Similar structures had significantly greater enhancement when 120 kV was applied instead of 140 kV (P =.0495)
CONCLUSION: The 16-detector row CT scanner improved gadolinium-enhanced CT angiography results compared with the four-detector row CT unit.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15361562     DOI: 10.1097/01.RVI.0000130381.73361.7A

Source DB:  PubMed          Journal:  J Vasc Interv Radiol        ISSN: 1051-0443            Impact factor:   3.464


  6 in total

1.  Scan-timed CT angiography in experimental arthritis: selective placement of regions of interest.

Authors:  Stephanie Holowka; Andrea S Doria
Journal:  Pediatr Radiol       Date:  2005-09-15

2.  Dual-contrast agent photon-counting computed tomography of the heart: initial experience.

Authors:  Rolf Symons; Tyler E Cork; Manu N Lakshmanan; Robert Evers; Cynthia Davies-Venn; Kelly A Rice; Marvin L Thomas; Chia-Ying Liu; Steffen Kappler; Stefan Ulzheimer; Veit Sandfort; David A Bluemke; Amir Pourmorteza
Journal:  Int J Cardiovasc Imaging       Date:  2017-03-13       Impact factor: 2.357

3.  Feasibility of gadolinium-enhanced dual energy CT pulmonary angiography: a pilot study in rabbits.

Authors:  Long Jiang Zhang; Zhen Jane Wang; Li Lu; Bo Chen; Guang Ming Lu
Journal:  Int J Cardiovasc Imaging       Date:  2010-11-26       Impact factor: 2.357

Review 4.  Clinical insights into the diagnosis and management of atherosclerotic renal artery disease.

Authors:  Michael J Bloch; Jan Basile
Journal:  Curr Atheroscler Rep       Date:  2006-09       Impact factor: 5.113

5.  Iodine contrast iso-attenuating with diagnostic gadolinium doses in CTA and angiography results in ultra-low iodine doses. A way to avoid both CIN and NSF in azotemic patients?

Authors:  Ulf Nyman; Barbara Elmståhl; Håkan Geijer; Peter Leander; Torsten Almén; Mats Nilsson
Journal:  Eur Radiol       Date:  2010-08-29       Impact factor: 5.315

6.  One-molar gadolinium chelate (gadobutrol) as a contrast agent for CT angiography of the thoracic and abdominal aorta.

Authors:  José M Esteban; Antonio Alonso; Vicente Cervera; Vicente Martínez
Journal:  Eur Radiol       Date:  2007-02-07       Impact factor: 5.315

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.